R&S®FSC Spectrum Analyzer Specifications Data Sheet | Version 03.00 year FSC_dat-sw_en_5214-3330-22_v0300_cover.indd 1 01.02.
Version 03.00, February 2017 CONTENTS Base unit .......................................................................................................................................................................... 3 Frequency ......................................................................................................................................................................................... 3 Sweep time..................................................................................
Version 03.00, February 2017 Base unit Frequency Frequency range model .03/.13 model .06/.
Version 03.00, February 2017 Level Display range displayed noise floor to +30 dBm Maximum rated input level with RF attenuation ≥ 10 dB DC voltage 50 V CW RF power 30 dBm (= 1 W) Peak RF power < 3 s duration 33 dBm (= 2 W) Max. pulse voltage 150 V Max. pulse energy pulse width 10 µs 10 mWs Maximum rated input level with RF attenuation < 10 dB DC voltage 50 V CW RF power 20 dBm (= 100 mW) Peak RF power < 3 s duration 23 dBm (= 200 mW) Max. pulse voltage 50 V Max.
Version 03.00, February 2017 Immunity to interference, nominal values Image frequencies fin – 2 × 21.4 MHz fin – 2 × 831.4 MHz fin – 2 × 4881 MHz Intermediate frequencies 21.4 MHz, 831.4 MHz, 4881.4 MHz 8931.4 MHz Other interfering signals, f ≤ 3.6 GHz signal level – RF attenuation < –20 dBm spurious at fin – 2440.7 MHz 3.6 GHz < f ≤ 6 GHz spurious at fin – 4465.7 MHz Other interfering signals, f ≤ 3.6 GHz related to local oscillators ∆f < 300 kHz ∆f ≥ 300 kHz f > 3.
Version 03.00, February 2017 Tracking generator (model .13/.16 only) Frequency range Connector VSWR Output level Tracking generator attenuator Dynamic range model .13 model .16 100 kHz to 3 GHz 100 kHz to 6 GHz N female, 50 Ω 100 kHz ≤ f ≤ 1 GHz < 1.5, nominal 1 GHz < f ≤ 3 GHz < 2, nominal 3 GHz < f ≤ 6 GHz (model .
Version 03.
Version 03.00, February 2017 General data Power supply AC supply DC supply Power consumption Safety input specifications input specifications Test mark Manual operation Languages Remote control Command set LAN interface USB interface Display Type Resolution Audio Speaker Mass memory Mass memory Data storage Temperature Climatic loading Mechanical resistance Vibration 100 V AC to 240 V AC, 50 Hz to 60 Hz, 400 Hz, 130 VA 14 V to 16 V, 0.9 A to 0.
Version 03.00, February 2017 Ordering information Designation Type Order No. Spectrum Analyzer, 9 kHz to 3 GHz R&S®FSC3 1314.3006.03 Spectrum Analyzer, 9 kHz to 3 GHz, R&S®FSC3 1314.3006.13 with tracking generator Spectrum Analyzer, 9 kHz to 6 GHz R&S®FSC6 1314.3006.06 Spectrum Analyzer, 9 kHz to 6 GHz, R&S®FSC6 1314.3006.
Version 03.00, February 2017 Recommended extras Designation Ethernet Cable Headphones 19” Rack Adapter for installing two R&S®FSC 19” Rack Adapter for installing one R&S®FSC Matching pad 50/75 Ω, 0 Hz to 2700 MHz, matching at both ends, N-connectors Matching pad 50/75 Ω, 0 Hz to 2700 MHz, matching at one end, N-connectors 75 ohm matching pad N to BNC (female) Near-Field Probe Set Preamplifier for R&S®HZ-15 10 Type R&S®HA-Z210 R&S®FSH-Z36 R&S®ZZA-T33 Order No. 1309.6152.00 1145.5838.02 1109.4458.
Version 03.00, February 2017 Supported Power Sensors 3 Designation Type Universal Power Sensor, 10 MHz to 8 GHz, 100 mW, 2-path R&S®NRP-Z211 Universal Power Sensor, 10 MHz to 18 GHz, 100 mW, 2-path R&S®NRP-Z221 R&S®NRP-Zxx power sensors require the following adapter cable for operation on the R&S®FSC USB Adapter Cable (passive), length: 2 m (78.
Service that adds value Worldwide Local and personalized ❙ Customized and flexible ❙ Uncompromising quality ❙ Long-term dependability ❙ ❙ Rohde & Schwarz The Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and measurement, broadcast and media, secure communications, cybersecurity, monitoring and network testing.