User Manual

Operability outside the secured area
R&S
®
FPL1000
8Instrument Security Procedures 1178.6810.02 ─ 05
Figure 5-1: Location of the PC board for R&S
FPL1000
3. Unscrew the two knurled screws.
4. Remove the solid-state drive from the R&S FPL1000.
5. Keep the solid-state drive under organizational control.
6 Operability outside the secured area
As the solid-state drive holds the operating system, the R&S FPL1000 cannot be oper-
ated without the solid-state drive.
For servicing and calibration, Rohde & Schwarz provides a separate solid-state drive
(option R&S FPL1-B19). This solid-state drive contains the operating system and
required instrument data.
To use the instrument outside the secured area:
1. Insert the second solid-state drive.
This solid-state drive enables the R&S FPL1000 to start the operating system.
2. Turn on the R&S FPL1000.
3. Perform a self-alignment as described in
Chapter 7, "Validity of instrument calibra-
tion after sanitization", on page 9.
The instrument is ready for use.