User Manual
Remote Control
R&S
®
FPC
314User Manual 1178.4130.02 ─ 05
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Unwrap phase
DISP:PHAS:UNWR ON
DISPlay<n>:PHASe:Y:SCALe <DisplayRange>
This command defines the range of the y-axis for the phase measurement format.
Prerequisites for this command
●
Select reflection measurement (MEASurement<n>:MODE).
●
Select phase format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayRange> <numeric value>
The number you enter is rounded up to the next possible display
range. For example, if you enter 300°, the R&S FPC automati-
cally sets the display range to 360°.
Range: 90 to 100000
*RST: 360
Default unit: deg
Example: //Define display range for phase
MEAS:MODE REFL
MEAS:FORM PHAS
DISP:PHAS:Y:SCAL 180
DISPlay<n>:VSWR:Y:SCALe <DisplayRange>
This command defines the range of the y-axis for the SWR measurement format.
Prerequisites for this command
●
Select reflection or DTF measurement (MEASurement<n>:MODE).
●
Select SWR format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Remote Commands of the VNA Application