User Manual

Remote Control
R&S
®
FPC
313User Manual 1178.4130.02 ─ 05
DISPlay<n>:PHASe:REFerence <RefValue>
This command defines the reference value for the phase measurement format.
Prerequisites for this command
Select reflection measurement (MEASurement<n>:MODE).
Select phase format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<RefValue> <numeric value>
Range: -100000 to 100000
*RST: 0
Default unit: deg
Example: //Define reference value for phase
MEAS:MODE REFL
MEAS:FORM PHAS
DISP:PHAS:REF 360
DISPlay<n>:PHASe:REFerence:POSition <RefPosition>
This command defines the position of the reference value for the phase measurement
format.
Each step shifts the reference position one grid line up or down.
Prerequisites for this command
Select reflection measurement (MEASurement<n>:MODE).
Select phase format (MEASurement<n>:FORMat).
Suffix:
<n>
.
irrelevant
Parameters:
<RefPosition> <numeric value> (integer only)
Range: 0 to 10
*RST: 8
Example: //Define 5th grid line as the reference position for phase format
MEAS:MODE REFL
MEAS:FORM PHAS
DISP:PHAS:REF:POS 5
DISPlay<n>:PHASe:UNWRap <State>
This command removes the restriction limiting the value range to +/- 180°.
Suffix:
<n>
.
irrelevant
Remote Commands of the VNA Application