User Manual
Remote Control
R&S
®
FPC
307User Manual 1178.4130.02 ─ 05
Parameters:
<CableModel> String containing the file name of the cable model.
Example: //Select cable model
MEAS:MODE DTF
CALC:DTF:CAB:PRES '5088-HLFR.CBLMOD'
CALCulate<n>:DTF:DISTance:STARt <StartDistance>
This command defines the starting point (distance) of the DTF measurement.
Prerequisites for this command
●
Select the DTF measurement (MEASurement<n>:MODE).
Suffix:
<n>
.
irrelevant
Parameters:
<StartDistance> <numeric value>
The unit depends on your selection (UNIT<n>:LENGth).
*RST: 0 m
Default unit: m or ft
Example: //Define start distance for DTF measurement
UNIT:LENG MET
MEAS:MODE DTF
CALC:DTF:DIST:STAR 10
CALCulate<n>:DTF:DISTance:STOP <StopDistance>
This command defines the end point (distance) of the DTF measurement.
Prerequisites for this command
●
Select the DTF measurement (MEASurement<n>:MODE).
Suffix:
<n>
.
irrelevant
Parameters:
<StopDistance> <numeric value>
The unit depends on your selection (UNIT<n>:LENGth).
*RST: 50 m
Default unit: m or ft
Example: //Define stop distance for DTF measurement
UNIT:LENG MET
MEAS:MODE DTF
CALC:DTF:DIST:STOP 100
Remote Commands of the VNA Application