301User Manual 1178.4130.02 ─ 05
Example: //Define roll-off factor
DDEM:FILT:ALPH 0.35
[SENSe:]DDEMod:FILTer:NAME <FilterType>
This command selects the measurement filter.
You can define the filter shape with [SENSe:]DDEMod:FILTer:ALPHa.
Parameters:
<FilterType> If you want to use no filter (unfiltered measurements), use
[SENSe:]DDEMod:FILTer[:STATe].
GAUS
Select a Gaussian filter.
RC
Select a raised cosine filter.
RRC
Selects a root raised cosine filter.
*RST: GAUS
Example: //Select RRC filter
DDEM:FILT:NAME RRC
[SENSe:]DDEMod:FILTer[:STATe] <State>
This command turns the measurement filter on and off.
Parameters:
<State> OFF | 0
Turns off the measurement filter for unfiltered measurements.
ON | 1
Turns on the measurement filter selected with [SENSe:
]DDEMod:FILTer:NAME.
*RST: ON
Example: //Perform unfiltered measurement
DDEM:FILT OFF
[SENSe:]DDEMod:SRATe <SymbolRate>
This command defines the symbol rate.
Parameters:
<SymbolRate> <numeric value> (integer only)
Default unit: Hz
Example: //Define symbol rate
DDEM:SRAT 10KHZ
Remote Commands of the Digital Demodulation