User Manual

Remote Control
R&S
®
FPC
283User Manual 1178.4130.02 ─ 05
Example: //Select CISPT bandwidth
BAND:CISP 9KHZ
[SENSe:]BANDwidth[:RESolution]:CISPr:AUTO <State>
This command turns automatic selection of the CISPR measurement bandwidth on
and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Select CISPR bandwidth manually
BAND:CISP:AUTO OFF
BAND:CISP 120KHZ
17.13.5 Sweep Control
Commands to control the measurement described elsewhere.
Measurement control:
INITiate[:IMMediate]
Measurement type (single / continuous):
INITiate:CONTinuous
Measurement time:
[SENSe:]SWEep:TIME
17.13.6 Trigger
Commands to configure trigger described elsewhere.
Trigger slope:
TRIGger[:SEQuence]:SLOPe
TRIGger[:SEQuence]:SOURce........................................................................................283
TRIGger[:SEQuence]:SOURce <TriggerSource>
This command selects the trigger source.
Parameters:
<TriggerSource> IMMediate
No trigger is applied.
EXTernal
Select an external trigger.
*RST: IMMediate
Example: //Select external trigger
TRIG:SOUR EXT