User Manual
Remote Control
R&S
®
FPC
280User Manual 1178.4130.02 ─ 05
TRACe[:DATA]?
This command queries the trace data (measurement results in graphical result dis-
plays).
Prerequisites for this command
●
Select graphical result display (CALCulate<n>:FEED).
Return values:
<Results> Returns one value for each measurement point.
The type of value depends on the measurement (Hz for FM,
PCT for AM).
Example: //Query FM results
CALC:FEED 'XTIMe:FM'
TRAC?
Usage: Query only
17.13.2
Frequency
Commands to configure the frequency described elsewhere.
●
Frequency stepsize:
[SENSe:]FREQuency:CENTer:STEP
●
Frequency mode (frequency / channel):
[SENSe:]FREQuency:INPut:MODE
●
Channel selection:
[SENSe:]CHANnel
[SENSe:]FREQuency:CENTer......................................................................................... 280
[SENSe:]SCAN:STARt....................................................................................................281
[SENSe:]SCAN:STEP.....................................................................................................281
[SENSe:]SCAN:STOP.................................................................................................... 281
[SENSe:]FREQuency:CENTer <Frequency>
This command defines the receiver frequency.
Prerequisites for this command
●
Select fixed frequency scan ([SENSe:]FREQuency:MODE).
Parameters:
<Frequency> <numeric value>
Default unit: Hz
Example: //Define a receiver frequency for fixed frequency measurement
FREQ:MODE FIX
FREQ:CENT 100MHZ
Remote Commands of the Receiver Application