User Manual

Remote Control
R&S
®
FPC
279User Manual 1178.4130.02 ─ 05
Parameters:
<Level> <numeric value>
Unit depends on the unit you have currently selected for the ver-
tical axis.
Example: //Turn on threshold limit line
CALC:LIM:UPP:THR -10DBM
17.13 Remote Commands of the Receiver Application
Measurement Configuration..................................................................................279
Frequency............................................................................................................. 280
Amplitude.............................................................................................................. 282
Bandwidth............................................................................................................. 282
Sweep Control.......................................................................................................283
Trigger...................................................................................................................283
Traces................................................................................................................... 284
Marker...................................................................................................................285
17.13.1 Measurement Configuration
DISPlay:BARGraph:PHOLd:RESet.................................................................................. 279
[SENSe:]FREQuency:MODE...........................................................................................279
TRACe[:DATA]?.............................................................................................................280
DISPlay:BARGraph:PHOLd:RESet
This command resets the maxhold information in the fixed frequency measurement
and the frequency scan.
Example:
//Reset maxhold information
DISP:BARG:PHOL:RES
Usage: Event
[SENSe:]FREQuency:MODE <Mode>
This command select the receiver measurement mode.
Parameters:
<Mode> FIXed | CW
Selects fixed frequency measurements.
SCAN
Selects frequency scan measurements.
*RST: FIXed
Example: //Select frequency scan
FREQ:MODE SCAN
Remote Commands of the Receiver Application