User Manual
Remote Control
R&S
®
FPC
275User Manual 1178.4130.02 ─ 05
Example: //Minimum requirement of two data points
CALC:LIM:DEF 'Limit','Empty
comment',HZ,ABS,DBM,0,-70,100000000000,-70
CALC:LIM:UPP:SEL 'limit.abslim'
Example: //More complex limit line
CALC:LIM:DEF 'Limit','Empty
comment',HZ,ABS,DBM,
0,-70,100000000,-70,100000000,-80,1000000000,
-80,1000000000,-70,2000000000,-70
CALC:LIM:UPP:SEL 'limit.abslim'
Usage: Setting only
CALCulate<n>:LIMit<k>:DELete
This command removes a limit line from the evaluation.
Removing a limit line also ends the limit check.
Prerequisites for this command
●
Select a limit line or define a threshold limit with one of the following commands:
– CALCulate<n>:LIMit<k>:LOWer:SELect
– CALCulate<n>:LIMit<k>:LOWer:THReshold
– CALCulate<n>:LIMit<k>:UPPer:SELect
– CALCulate<n>:LIMit<k>:UPPer:THReshold
Suffix:
<n>
.
irrelevant
<k> Limit line
Example: //Remove limit line
CALC:LIM2:DEL
CALCulate<n>:LIMit<k>:FAIL?
This command queries the result of a limit check.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end before reading out the result. This is only possible in single sweep mode.
Prerequisites for this command
●
Select a limit line or define a threshold limit with one of the following commands:
– CALCulate<n>:LIMit<k>:LOWer:SELect
– CALCulate<n>:LIMit<k>:LOWer:THReshold
– CALCulate<n>:LIMit<k>:UPPer:SELect
– CALCulate<n>:LIMit<k>:UPPer:THReshold
Suffix:
<n>
.
irrelevant
Remote Commands of the Spectrum Application