272User Manual 1178.4130.02 ─ 05
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe] <State>
This command turns the noise measurement marker function on and off.
You can query the results of the noise power density at the marker position with
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
Suffix:
<n>
.
irrelevant
<m> Marker
(Only marker 1 (M1) or delta marker 2 (D2) are possible to
select.)
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0
Example: See CALCulate<n>:MARKer<m>:FUNCtion:NOISe:
RESult?
17.12.10 Display Lines
CALCulate<n>:DLINe.....................................................................................................272
CALCulate<n>:DLINe:STATe...........................................................................................272
CALCulate<n>:DLINe <LinePosition>
This command defines the position of a display line.
Prerequisites for this command
●
Turn on a display line (CALCulate<n>:DLINe:STATe).
Suffix:
<n>
.
irrelevant
Parameters:
<LinePosition> <numeric value>
You can use any unit you want, the R&S FPC then converts the
unit to the currently selected unit. If you omit a unit, the
R&S FPC uses the currently selected unit.
*RST: - (STATe to OFF)
Default unit: -
Example: //Position display line
CALC:DLIN:STAT ON
CALC:DLIN –20dBm
CALCulate<n>:DLINe:STATe <State>
This command turns display lines on and off.
Remote Commands of the Spectrum Application