User Manual

Remote Control
R&S
®
FPC
269User Manual 1178.4130.02 ─ 05
Parameters:
<Demodulation> AM | FM
*RST: -
Example: See CALCulate<n>:MARKer<m>:FUNCtion:
DEModulation:HOLD.
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe] <State>
This command turns marker demodulation on and off.
Prerequisites for this command
Turn on a marker (CALCulate<n>:MARKer<m>[:STATe]).
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0
Example: See CALCulate<n>:MARKer<m>:FUNCtion:
DEModulation:HOLD.
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown <MarkerDistance>
This command defines the distance of the n dB down markers to the reference marker.
Prerequisites for this command
Turn on n dB down measurement (CALCulate<n>:MARKer<m>:FUNCtion:
NDBDown:STATe).
Suffix:
<n>
.
irrelevant
<m> irrelevant (the n dB down marker function always uses marker 1)
Parameters:
<MarkerDistance> <numeric value>
Distance of the temporary markers to the reference marker in
dB.
*RST: 3 dB
Default unit: dB
Example: See CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:
STATe
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
This command queries the position of the n dB down markers on the x-axis.
Remote Commands of the Spectrum Application