CALCulate<n>:MARKer<m>:FUNCtion:NDBDown............................................................ 269
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?........................................269
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult?............................................... 270
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:STATe.................................................. 270
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?.....................................................271
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]......................................................272
CALCulate<n>:MARKer<m>:COUNt:FREQuency?
This command measures the frequency at the marker position and returns the result.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S FPC actually reaches the frequency you want to
measure. This is only possible in single sweep mode.
Prerequisites for this command
●
Turn on the frequency counter (CALCulate<n>:MARKer<m>:COUNt[:STATe]).
Suffix:
<n>
.
irrelevant
<m> irrelevant
Return values:
<Frequency> <numeric value>
Default unit: Hz
Example: See CALCulate<n>:MARKer<m>:COUNt[:STATe]
Usage: Query only
CALCulate<n>:MARKer<m>:COUNt[:STATe] <State>
This command turns the frequency counter at the marker position on and off.
You can read out the result with CALCulate<n>:MARKer<m>:COUNt:FREQuency?
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S FPC actually reaches the frequency you want to
measure. This is only possible in single sweep mode.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0