User Manual

Remote Control
R&S
®
FPC
262User Manual 1178.4130.02 ─ 05
CALCulate<n>:MARKer<m>:AOFF
This command turns off all active markers, delta markers and active marker measure-
ment functions.
This command is an event and therefore has no query and no *RST value.
Suffix:
<n>
.
irrelevant
<m> Marker
Example: //Turn off all markers
CALC:MARK:AOFF
Usage: Event
CALCulate<n>:MARKer<m>:FREQuency:MODE
<Mode>
This command selects the marker frequency display mode.
Suffix:
<n>
.
irrelevant
<m> Marker
Parameters:
<Mode> FREQuency
Set the marker frequency mode to frequency input (in Hz).
CHANnel
Set the marker frequency mode to channel input (as a channel
number).
Example: //Select the frequency display mode.
CALC:MARK:FREQ:MODE FREQ
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
Suffix:
<n>
.
irrelevant
<m> Marker
Example: //Match the center frequency to the frequency of marker 1
CALC:MARK1:FUNC:CENT
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
Suffix:
<n>
.
irrelevant
Remote Commands of the Spectrum Application