257User Manual 1178.4130.02 ─ 05
[SENSe:]DETector<t>[:FUNCtion]:AUTO <State>
This command couples and decouples the detector to the trace mode.
Suffix:
<t>
.
Trace
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON | 1
Example: //Turn off automatic detector selection
DET:AUTO OFF
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of averages for the average trace mode.
The R&S FPC performs one sweep for sweep count 0 or 1.
Prerequisites for this command
●
Select average trace mode (DISPlay<n>[:WINDow]:TRACe<t>:MODE).
Parameters:
<SweepCount> <numeric value> (integer)
Range: 0 to 999
*RST: 1
Example: //Average over 64 sweeps
DISP:TRAC:MODE AVER
SWE:COUN 64
//Turn on single sweep mode, start the sweep and wait for its
end
INIT:CONT OFF
INIT;*WAI
TRACe<n>[:DATA]? <ResultType>
This command queries the trace data of the current measurement.
It also transfers data from a file to a certain trace.
With the command FORMat[:DATA], you can set the data format.
Suffix:
<n>
.
irrelevant
Query parameters:
<ResultType> TRACE1
Query the data of trace 1.
TRACE2
Query the data of trace 2.
Remote Commands of the Spectrum Application