Example: //Abort measurement and start a new one once abort is com-
plete
ABOR;*WAI
INIT:IMM
Usage: Event
[SENSe:]SWEep:TIME <SweepTime>
This command defines the sweep time.
Effects of this command
Turns off automatic selection of the sweep time ([SENSe:]BANDwidth[:
RESolution]:AUTO).
Parameters:
<SweepTime> <numeric value>
The available value range is specified in the datasheet.
*RST: - (AUTO is set to ON)
Default unit: s
Example: //Define sweep time
SWE:TIME 10s
[SENSe:]SWEep:TIME:AUTO <State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Effects of this command
Turn on automatic sweep time selection to couple the sweep time to the span and
the RBW.
When you turn off automatic selection of the sweep time, you can select one manually
with [SENSe:]SWEep:TIME.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON | 1
Example: //Turn on automatic selection of the sweep time
SWE:TIME:AUTO ON
INITiate:CONTinuous <State>
This command selects the sweep mode.
To actually start the measurement, use INITiate[:IMMediate].