User Manual

Remote Control
R&S
®
FPC
249User Manual 1178.4130.02 ─ 05
Example: //Select video bandwidth
BAND:VID 10KHZ
[SENSe:]BWIDth:VIDeo:AUTO <State>
[SENSe:]BANDwidth:VIDeo:AUTO <State>
This command turns automatic selection of the resolution bandwidth on and off.
Effects of this command
Turn on automatic VBW selection to couple the VBW to the RBW.
When you turn off automatic selection of the video bandwidth, you can select one man-
ually with [SENSe:]BANDwidth:VIDeo.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON | 1
Example: //Select VBW manually
BAND:VID:AUTO OFF
BAND:VID 10HZ
17.12.5 Sweep Control
ABORt.......................................................................................................................... 249
[SENSe:]SWEep:TIME................................................................................................... 250
[SENSe:]SWEep:TIME:AUTO......................................................................................... 250
INITiate:CONTinuous......................................................................................................250
INITiate[:IMMediate]....................................................................................................... 251
ABORt
This command aborts a measurement and resets the trigger system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the *OPC? or *WAI command after ABORt
and before the next command.
Note on blocked remote control programs
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S FPC is blocked for further commands. In this case, you must inter-
rupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the R&S FPC
on a parallel channel to clear all currently active remote channels. Depending on the
used interface and protocol, send the following commands
Visa:viClear()
Now you can send the ABORt command on the remote channel performing the mea-
surement.
Remote Commands of the Spectrum Application