User Manual

Remote Control
R&S
®
FPC
248User Manual 1178.4130.02 ─ 05
Effects of this command
Turns off automatic selection of resolution bandwidth ([SENSe:]BANDwidth[:
RESolution]:AUTO).
Parameters:
<Bandwidth> <numeric value> (integer)
Value adjustable in 1 - 3 steps (1 Hz, 3 Hz, 10 Hz, etc.).
In addition, you can select a 200 kHz bandwidth.
Range: 1 Hz to 3 MHz
*RST: - (AUTO is set to ON)
Default unit: Hz
Example: //Define resolution bandwidth
BAND 100KHZ
[SENSe:]BWIDth[:RESolution]:AUTO
<Bandwidth>
[SENSe:]BANDwidth[:RESolution]:AUTO <State>
This command turns automatic selection of the resolution bandwidth (or measurement
bandwidth) on and off.
Effects of this command
Turn on automatic RBW selection to couple the RBW to the span.
When you turn off automatic selection of the resolution bandwidth, you can select one
manually with [SENSe:]BANDwidth[:RESolution].
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON | 1
Example: //Turn on automatic RBW selection
BAND:AUTO OFF
[SENSe:]BWIDth:VIDeo <Bandwidth>
[SENSe:]BANDwidth:VIDeo <Bandwidth>
This command selects the video bandwidth.
Effects of this command
Turns off automatic selection of video bandwidth ([SENSe:]BANDwidth:VIDeo:
AUTO).
Parameters:
<Bandwidth> <numeric value> (integer)
Value adjustable in 1 - 3 steps (1 Hz, 3 Hz, 10 Hz, etc.).
Range: 1 Hz to 3 MHz
*RST: - (AUTO is set to ON)
Default unit: Hz
Remote Commands of the Spectrum Application