User Manual

Remote Control
R&S
®
FPC
243User Manual 1178.4130.02 ─ 05
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel <ReferenceLevel>
This command defines the reference level.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Parameters:
<ReferenceLevel> <numeric value>
The unit depends on UNIT<n>:POWer.
The available value range is specified in the data sheet.
*RST: -20 dBm
Example: //Define reference level
UNIT:POW DBM
DISP:TRAC:Y:RLEV –60
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet <Offset>
This command defines a reference level offset.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Parameters:
<Offset> <numeric value>
Range: -100 to 100 dB
*RST: 0 dB
Default unit: dB
Example: //Define a level offset
DISP:TRAC:Y:RLEV:OFFS –10dB
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RPOSition <Position>
This command defines the position of the reference level on the display grid.
Prerequisites for this command
Select logarithmic scale for the y-axis (DISPlay<n>[:WINDow]:TRACe<t>:Y:
SPACing).
For a linear scale, you cannot modify the reference position.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Remote Commands of the Spectrum Application