User Manual

Remote Control
R&S
®
FPC
242User Manual 1178.4130.02 ─ 05
Parameters:
<Scaling> LOGarithmic
Select a logarithmic scale.
When you select a logarithmic scale, you can define the display
range manually or automatically.
Manual: DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]
Automatic: DISPlay<n>[:WINDow]:TRACe<t>:Y[:
SCALe]:ADJust
LINear
Select a linear scale (%).
The display range of a linear scale is fixed (0 % to 100 %).
*RST: LOGarithmic
Example: //Select linear scaling of the y-axis
DISP:TRAC:Y:SPAC LIN
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe] <DisplayRange>
This command defines the display range of the vertical axis.
Prerequisites for this command
Select logarithmic scale for the y-axis (DISPlay<n>[:WINDow]:TRACe<t>:Y:
SPACing).
For a linear scale, you cannot modify the display range as it is fixed.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Parameters:
<DisplayRange> <numeric value>
Range: 1 to 150 dB
*RST: 100 dB
Default unit: dB
Example: //Define display range
DISP:TRAC:Y:SPAC LOG
DISP:TRAC:Y 110dB
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:ADJust
This command automatically scales the y-axis.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Example: //Adjust the y-axis
DISP:TRAC:Y:ADJ
Usage: Event
Remote Commands of the Spectrum Application