User Manual
Remote Control
R&S
®
FPC
241User Manual 1178.4130.02 ─ 05
Example: //Define start frequency
FREQ:STAR 20MHZ
[SENSe:]FREQuency:STOP <Frequency>
This command defines the stop frequency for measurements in the frequency domain
(span > 0).
Parameters:
<Frequency> <numeric value>
Refer to the datasheet for details on the frequency range of your
R&S FPC.
*RST: fmax
Default unit: Hz
Example: //Define stop frequency
FREQ:STOP 500MHZ
17.12.3 Amplitude
DISPlay<n>[:WINDow]:TRACe<t>:Y:SPACing...................................................................241
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]................................................................... 242
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:ADJust........................................................242
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel........................................................243
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet........................................... 243
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RPOSition................................................... 243
INPut:ATTenuation......................................................................................................... 244
INPut:ATTenuation:AUTO............................................................................................... 244
INPut:ATTenuation:MODE...............................................................................................244
INPut:GAIN:STATe......................................................................................................... 245
INPut:IMPedance...........................................................................................................245
INPut:IMPedance:PAD....................................................................................................245
[SENSe:]CORRection:TRANsducer<td>:SELect............................................................... 246
[SENSe:]CORRection:TRANsducer<td>:UNIT?.................................................................246
[SENSe:]CORRection:TRANsducer<td>[:STATe]...............................................................246
UNIT<n>:POWer............................................................................................................247
DISPlay<n>[:WINDow]:TRACe<t>:Y:SPACing <Scaling>
This command selects the scale of the y-axis.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Remote Commands of the Spectrum Application