User Manual
Remote Control
R&S
®
FPC
235User Manual 1178.4130.02 ─ 05
CALCulate<n>:DELTamarker<m>:X:RELative
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult?...................................................235
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]................................................... 235
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult?
This command queries the results of the AM modulation depth measurement.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Prerequisites for this command
●
Turn on modulation depth measurement (CALCulate<n>:MARKer<m>:
FUNCtion:MDEPth[:STATe]).
Suffix:
<n>, <m>
.
irrelevant
Return values:
<ModulationDepth> <numeric value>
Modulation depth in %.
Example: //Query the AM modulation depth.
CALC:MARK:FUNC:MDEP:RES?
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe] <State>
This command turns the AM modulation depth measurement on and off.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State> ON | 1
Turns on the AM modulation depth measurement.
OFF | 0
Performs a basic spectrum measurement.
*RST: OFF
Example: //Turn on AM modulation depth measurement
CALC:MARK:FUNC:MDEP ON
17.12.2 Frequency
[SENSe:]CHANnel......................................................................................................... 236
[SENSe:]CHANnel:TABLe:SELect:DOWNlink....................................................................236
[SENSe:]CHANnel:TABLe:SELect:UPLink........................................................................ 236
[SENSe:]CHANnel:TABLe:SELect....................................................................................237
Remote Commands of the Spectrum Application