User Manual

Remote Control
R&S
®
FPC
235User Manual 1178.4130.02 ─ 05
CALCulate<n>:DELTamarker<m>:X:RELative
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult?...................................................235
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]................................................... 235
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult?
This command queries the results of the AM modulation depth measurement.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Prerequisites for this command
Turn on modulation depth measurement (CALCulate<n>:MARKer<m>:
FUNCtion:MDEPth[:STATe]).
Suffix:
<n>, <m>
.
irrelevant
Return values:
<ModulationDepth> <numeric value>
Modulation depth in %.
Example: //Query the AM modulation depth.
CALC:MARK:FUNC:MDEP:RES?
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe] <State>
This command turns the AM modulation depth measurement on and off.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State> ON | 1
Turns on the AM modulation depth measurement.
OFF | 0
Performs a basic spectrum measurement.
*RST: OFF
Example: //Turn on AM modulation depth measurement
CALC:MARK:FUNC:MDEP ON
17.12.2 Frequency
[SENSe:]CHANnel......................................................................................................... 236
[SENSe:]CHANnel:TABLe:SELect:DOWNlink....................................................................236
[SENSe:]CHANnel:TABLe:SELect:UPLink........................................................................ 236
[SENSe:]CHANnel:TABLe:SELect....................................................................................237
Remote Commands of the Spectrum Application