User Manual
Remote Control
R&S
®
FPC
230User Manual 1178.4130.02 ─ 05
Suffix: 
<n>, <m>
.
irrelevant
Setting parameters: 
<Once> ONCE
Example:  //Search for signals
CALC:MARK:FUNC:TOI ON
CALC:MARK:FUNC:TOI:SEAR
Usage:  Setting only
CALCulate<n>:MARKer<n>:FUNCtion:TOI[:STATe] 
<State>
This command turns the third order intermodulation measurement on and off.
Suffix: 
<n>, <m>
.
irrelevant
Parameters:
<State> ON | 1
Turns on the third order intermodulation measurement.
OFF | 0
Performs a basic spectrum measurement.
*RST:  OFF
Example:  //Measure the TOI
CALC:MARK:FUNC:TOI ON
17.12.1.4  Occupied Bandwidth
The following remote commands are exclusive to the occupied bandwidth measure-
ment.
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth.................................................. 230
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth.................................................230
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth:PCT...........................................231
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth:PCT......................................... 231
CALCulate<n>:MARKer<m>:FUNCtion:OBW:BANDwidth <ChannelBW>
CALCulate<n>:MARKer<m>:FUNCtion:OBAN:BANDwidth <ChannelBW>
This command defines the channel bandwidth for occupied bandwidth measurements.
Prerequisites for this command
●
Turn on occupied bandwidth measurement (CALCulate<n>:MARKer<m>:
FUNCtion:POWer:SELect).
Suffix: 
<n>, <m>
.
irrelevant
Remote Commands of the Spectrum Application