User Manual
Remote Control
R&S
®
FPC
228User Manual 1178.4130.02 ─ 05
Parameters:
<Bandwidth> <numeric value>
*RST: 3.84 MHz
Default unit: Hz
Example: //Define the channel bandwidth
CALC:MARK:FUNC:CPOW:BAND 10MHZ
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:MODE <DisplayMode>
This command selects the display mode for channel power measurements.
Prerequisites for this command
●
Select the channel power measurement (CALCulate<n>:MARKer<m>:
FUNCtion:POWer:SELect).
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<DisplayMode> CLR
Display the currently measured value ("Clear / Write").
MAX
Display the highest measured value ("Max Hold").
*RST: CLR
Example: //Select clear/write trace mode
CALC:MARK:FUNC:CPOW:MODE CLR
CALCulate<n>:MARKer<m>:FUNCtion:CPOWer:UNIT <Unit>
This command selects the unit of the for channel power measurements.
Prerequisites for this command
●
Select the channel power measurement (CALCulate<n>:MARKer<m>:
FUNCtion:POWer:SELect).
Suffix:
<n>
.
irrelevant
<m> Marker
Parameters:
<Unit> DBM | DBMV | DBUV | VOLT | WATT | V | W
*RST: DBM
Example: //Select the unit for the channel power
CALC:MARK:FUNC:CPOW:UNIT DBM
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ <State>
This command turns the display of the channel power in a 1 Hz bandwidth on and off.
Remote Commands of the Spectrum Application