User Manual

Remote Control
R&S
®
FPC
221User Manual 1178.4130.02 ─ 05
'Connect Through' Connect a through to ports 1 and 2.
'Disconnect DUT from port 1'
'Calibration Done' Calibration has been finished. No further actions required.
CALibration<n>:MODE?
This command queries the current calibration type.
Suffix:
<n>
.
irrelevant
Return values:
<State> 0
Not calibrated or default calibration.
1
User calibration.
Example: //Query calibration type
CAL:MODE?
Usage: Query only
CALibration<n>:STARt? <CalibrationMethod>
This command selects a calibration method and initiates calibration.
After the command has been sent, it returns instructions on what to do to continue the
calibration process and finish it successfully.
Suffix:
<n>
.
irrelevant
Parameters for setting and query:
<CalibrationMethod> S11Cal
Full 1-port calibration.
Unavailable for S21 transmission measurements.
S11easy
Easy 1-port calibration.
Unavailable for S21 transmission measurements.
S21Norm
Normalization for S21.
Available for S21 transmission measurements.
Return values:
<Instructions> String containing the instructions for the next calibration step.
You can find a list possible instructions in the table below.
Calibration