User Manual

Remote Control
R&S
®
FPC
220User Manual 1178.4130.02 ─ 05
Example: //Turn on signal source
SOUR:TG ON
17.10 Calibration
CALibration<n>:ABORt...................................................................................................220
CALibration<n>:CONTinue?............................................................................................220
CALibration<n>:MODE?................................................................................................. 221
CALibration<n>:STARt?..................................................................................................221
CALibration<n>:STATus?................................................................................................222
SYSTem:PRESet:CALibration:DISCard............................................................................ 222
CALibration<n>:ABORt
This command aborts calibration.
Suffix:
<n>
.
irrelevant
Example: //Abort calibration
CAL:STAR? S11C
'Connect Open to Port 1'
CAL:ABOR
Usage: Event
CALibration<n>:CONTinue?
This command resumes calibration after a calibration standard has been connected.
Suffix:
<n>
.
irrelevant
Return values:
<Instructions> String containing the instructions for the next calibration step.
You can find a list possible instructions in the table below.
Example: //Resume calibration
CAL:STAR? S11C
'Connect Open to Port 1'
//Connect Open
CAL:CONT?
'Connect Short to Port 1'
Usage: Query only
Table 17-6: Return values: Possible instructions for calibration
'Connect Open to Port 1' Connect the open to port 1 before resuming calibration.
'Connect Short to Port 1' Connect the short to port 1 before resuming calibration.
'Connect Load to Port 1' Connect the load to port 1 before resuming calibration.
Calibration