Prerequisites for this command
Turn on signal source (SOURce:TG[:STATe]).
Turn off tracking generator (SOURce:TG:AUTO).
Parameters:
<State> ON | 1
Generates a signal whose frequency corresponds to the center
frequency (signal source mode "Coupled CW").
You can define the center frequency with [SENSe:
]FREQuency:CENTer.
OFF | 0
Generates a signal whose frequency remains the same (signal
source mode "CW").
You can define the fix output frequency with SOURce:TG:
FREQuency.
Example: //Couple signal source frequency to center frequency
SOUR:TG ON
SOUR:TG:FREQ:AUTO ON
//Define center and output frequency
FREQ:CENT 100MHZ
SOURce:TG:POWer <Level>
This command defines signal source output power.
Prerequisites for this command
Turn on signal source (SOURce:TG[:STATe]).
Parameters:
<Level> <numeric level>
*RST: 0
Default unit: dBm
Example: //Define signal source output power
SOUR:TG ON
SOUR:TG:POW -10DBM
SOURce:TG[:STATe] <State>
This command turns the signal source on and off.
Parameters:
<State> ON | OFF | 1 | 0
Configure the frequency characteristics of the output signal with
one of the following commands.
CW signal: SOURce:TG:FREQuency:AUTO
Tracking generator: SOURce:TG:AUTO
Configure the output level with SOURce:TG:POWer.
*RST: OFF
Signal Source Configuartion