Datasheet

Version 05.00, February 2019
4 Rohde & Schwarz R&S
®
FPC Spectrum Analyzer
Specifications
Frequency
Frequency range
R&S
®
FPC1000/R&S
®
FPC1500
5 kHz to 1 GHz
with R&S
®
FPC-B2 option installed
5 kHz to 2 GHz
with R&S
®
FPC-B2 and
R&S
®
FPC-B3 option installed
5 kHz to 3 GHz
Frequency resolution
1 Hz
Reference frequency, internal
Aging per year
1 × 10
6
Temperature drift
0 °C to +30 °C
1 × 10
6
+30 °C to +50 °C
3 × 10
6
Achievable initial calibration accuracy
5 × 10
7
Total reference uncertainty
(time since last adjustment × aging rate) +
temperature drift + calibration accuracy
Frequency readout
Marker resolution
0.1 Hz
Uncertainty
±(marker frequency × reference
uncertainty + 10 % × resolution bandwidth
+ ½ (span / (sweep points 1)) + 1 Hz)
Number of sweep (trace) points
1183
Marker tuning frequency step size
span/1182
Frequency counter resolution
0.1 Hz
Count uncertainty
SNR > 25 dB
±(frequency × reference uncertainty +
½ (last digit))
Frequency span
0 Hz, 10 Hz to 1 GHz
with R&S
®
FPC-B2 option installed
0 Hz, 10 Hz to 2 GHz
with R&S
®
FPC-B2 and
R&S
®
FPC-B3 option installed
0 Hz, 10 Hz to 3 GHz
Span uncertainty
1 % (nom.)
Spectral purity SSB phase noise
f = 500 MHz
Carrier offset
30 kHz
< 88 dBc (1 Hz), 92 dBc (1 Hz) (typ.)
100 kHz
< 98 dBc (1 Hz), 103 dBc (1 Hz) (typ.)
1 MHz
< 120 dBc (1 Hz), 125 dBc (1 Hz) (typ.)
Sweep time
Sweep time
100 µs to 100 s
20 ms to 1000 s
20 ms × span/600 MHz to 1000 s
Uncertainty
1 % (nom.)
3 % (nom.)
Bandwidth
Resolution bandwidths
Range
1 Hz to 3 MHz in 1/3 sequence
Bandwidth accuracy
< 5 % (nom.)
< 10 % (nom.)
Selectivity 60 dB:3 dB
< 5 (nom.) (Gaussian type filters)
Video filters
Range
1 Hz to 3 MHz in 1/3 sequence