Brochure

14
Specifications in brief
Frequency range R&S®FPC1000/R&S®FPC1500 base unit 5kHz to 1GHz
with R&S®FPC-B2 option 5kHz to 2GHz
with ¸FPC-B2 and R&S®FPC-B3 options 5kHz to 3GHz
Frequency resolution 1Hz
Resolution bandwidth 1Hz to 3MHz in 1/3 sequence
Spectral purity, SSB phase noise f = 500MHz
30kHz < –88 dBc (1Hz), typ. –92 dBc (1Hz)
100kHz < –98 dBc (1Hz), typ. –103 dBc (1Hz)
1MHz < –120 dBc (1Hz), typ. –125 dBc (1Hz)
Displayed average noise level 0 dB RF attenuation, termination 50 Ω, RBW = 100Hz, VBW = 10Hz,
sample detector, log scaling, normalized to 1Hz
preamplifier = off
1MHz to 10MHz < –127 dBm, typ. –135 dBm
10MHz to 2GHz < –142 dBm, typ. –150 dBm
2GHz to 3GHz < –138 dBm, typ. –147 dBm
preamplifier = on (requires R&S®FPC-B22 option)
1MHz to 10MHz < –147 dBm, typ. –157 dBm
10MHz to 2GHz < –158 dBm, typ. –165 dBm
2GHz to 3GHz < –155 dBm, typ. –163 dBm
Third-order intercept (TOI) intermodulation-free dynamic range, signal level 2 × –20 dBm, RF attenuation = 0 dB,
RF preamplifier = off
f
in
= 1GHz +7 dBm (meas.)
f
in
= 2.4GHz +10 dBm (meas.)
One-port vector network analyzer frequency range 2 MHz to 1/2/3 GHz
output power –10 dBm
Tracking generator frequency range 5 kHz to 1/2/3 GHz
output power –30 dBm to 0 dBm
Independent source frequency range 5 kHz to 1/2/3 GHz
output power –30 dBm to 0 dBm
Specifications in brief
FPC_bro_en_5214-7112-12_v0500.indd 14 31.01.2019 14:54:40