Datasheet
HMS3000/3010E/060912 · C&E · Subject to change without notice · © HAMEG Instruments GmbH
®
· DQS-certi fied in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · info@hameg.com
1 GHz Spectrum Analyzer HMS1000, HMS1010 (with TG)
[3 GHz Spectrum Analyzer HMS3000, HMS3010 (with TG)]
All data valid at 23 °C after 30 minute warm-up.
Frequency
Frequency range:
HMS1000, HMS1010
100 kHz…1 GHz
HMS3000, HMS3010
100 kHz…3 GHz
Temperature stability:
±2 ppm (0…30 °C)
Aging:
±1 ppm/year
Frequency counter:
Resolution
1 Hz
Accuracy
±(Frequency x tolerance of reference)
Span setting range:
HMS1000, HMS1010
0 Hz (zero span) and 100 Hz…1 GHz
HMS3000, HMS3010
0 Hz (zero span) and 100 Hz…3 GHz
Spectral purity, SSB phase noise:
30 kHz from carrier
(500 MHz, +20…30 °C)
<-85 dBc/Hz
100 kHz from carrier
(500M Hz, +20…30 °C)
<-100 dBc/Hz
1 MHz from carrier
(500MHz, +20…30 °C)
<-120 dBc/Hz
Sweep time:
Span = 0 Hz
2 ms…100 s
Span > 0 Hz
20 ms…1,000 s, min. 20 ms/600 MHz
Resolution bandwidths
(-3 dB):
100 Hz…1 MHz in 1–3 steps,
200 kHz
Tolerance
≤300 kHz
±5 % typ.
1 MHz
±10 % typ.
Resolution bandwidths
(-6 dB):
200 Hz, 9 kHz, 120 kHz, 1 MHz
Video bandwidths:
10 Hz…1 MHz in 1–3 steps
Amplitude
Display range:
Average noise level displayed up to
+2 0 d B m
Amplitude measurement
range:
Typ. -114…+20 dBm
Max. permissible DC
at HF input:
80 V
Max. power at HF input:
20 dBm, 30 dBm for max. 3 Min.
Intermodulation free range:
TOI products, 2 x -20 dBm
(-10 dBm ref. level)
66 dB typ.
(typ. +13 dBm third-order intercept)
(at distance between
signals ≤2 MHz)
60 dB typ. (+10 dBm TOI)
(at distance between
signals >2 MHz)
66 dB typ. (typ. +13 dBm TOI)
DANL (Displayed average noise level):
(RBW 100 Hz, VBW 10 Hz,
ref. level ≤-30 dBm
10 MHz…1 GHz resp. 3 GHz)
-115 dBm, typ. -124 dBm
With Preamp. -135 dBm typ.
Inherent spurious:
(ref. level ≤-20 dBm,
f >30 MHz, RBW ≤100 kHz)
<-80 dBm
Input related spurious:
(Mixer level ≤-40 dBm,
carrier offset >1 MHz)
-70 dBc typ., [-55 dBc (2…3 GHz)]
2
nd
harmonic receive frequency:
(mixer level -40 dBm) -60 dBc typ.
Level display:
Reference level
-80…+20 dBm in 1 dB steps
Display range
100 dB, 50 dB, 20 dB, 10 dB, linear
Logarithmic
display scaling
dBm, dBµV, dBmV
Linear display scaling
Percentage of reference level
Measured curves:
1 curve and 1 memory curve
Trace mathematics:
A-B (curve-stored curve), B-A
Detectors:
Auto-, Min-, Max-Peak, Sample, RMS,
Average, Quasi-Peak
Failure of level display:
<1.5 dB, typ. 0.5 dB
(ref. level -50 dBm, 20…30 °C)
Marker/Deltamarker
Number of marker:
8
Marker functions:
Peak, next peak, minimum,
center = marker, frequency,
reference level = marker level,
all marker on peak
Marker displays:
Normal (level, lin. & log.), delta marker,
noise marker, (frequency) counter
Inputs/Outputs
HF Input:
N socket
Input Impedance
50 Ω
VSWR
(10 MHz…1 GHz/3 GHz)
<1.5 typ.
Output tracking generator:
(HMS1010/HMS3010)
N socket
Output Impedance
50 Ω
Frequency range
5 MHz…1 GHz [3 GHz]
Output level
-20…0 dBm, in 1 dB steps
Trigger input:
BNC female
Trigger voltage
TTL
Ext. reference input/output:
BNC females
Reference frequency
10 MHz
Essential level (50 Ω)
10 dBm
Supply output for field
probes:
6 V
dc
, max. 100 mA (2.5 mm DIN jack)
Audio output (Phone):
3.5 mm DIN jack
Demodulation
AM and FM (internal speaker)
Miscellaneous
Display:
16.5 cm (6.5") TFT Color VGA Display
Save/Recall memory
10 complete device settings
Trigger
Free run, Video Trigger, Single Trigger,
external Trigger
Interfaces:
Dual-Interface USB/RS-232 (HO720),
USB-Stick (frontside),
USB-Printer (rear side),
DVI-D for ext. monitor
Power supply:
105…253 V, 50…60 Hz, CAT II
Power consumption:
Max. 40 W at 230 V, 50 Hz
Protection class:
Safety class I (EN61010-1)
Operating temperature:
+5…+40 °C
Storage temperature:
-20…+70 °C
Rel. humidity:
5…80 % (non condensing)
Dimensions (W x H x D):
285 x 175 x 220 mm
Weight:
3.6 kg
Accessories supplied: Line cord, Operating manual, HZ21 Adapter plug,
N-plug to BNC socket (2x HMS1010/3010), CD, Software
Recommended accessories:
HO730 Dual-Interface Ethernet/USB
HO740 Interface IEEE-488 (GPIB), galvanically isolated
HO3011 Preamplifier -135 dBm DANL (100 Hz RBW)
HZ13 Interface cable (USB) 1.8 m
HZ14 Interface cable (serial) 1:1
HZ20 Adapter, BNC to 4 mm banana
HZ33 Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34 Test cable 50 Ω, BNC/BNC, 1.0 m
HZ46 4RU 19" Rackmount Kit
HZ72 GPIB-Cable 2 m
HZ99 Carrying Case for protection and transport
HZ520 Plug-in Antenna with BNC connection
HZ525 50 Ω-Termination, N plug
HZ530 Near-Field Probe Set 1 GHz for EMI diagnostics
HZ540/550 Near-Field Probe Set 3 GHz for EMI diagnostics
HZ540L/550L Near-Field Probe Set 3 GHz for EMI diagnostics
HZ547 3 GHz VSWR Bridge for HMS1010, HMS3010
HZ560 Transient limiter
HZ575 75/50 Ω Converter
HZO30 Active probe 1 GHz (0.9 pF, 1 MΩ, including many accessories)