Mixed Signal Oscilloscope 50/70/100 MHz R&S®HMO1002
General information regarding the CE marking General information regarding the CE marking KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY DECLARATION DE CONFORMITE DECLARACIÓN DE CONFORMIDAD Hersteller / Manufacturer / Fabricant / Fabricante: HAMEG Instruments GmbH · Industriestraße 6 · D-63533 Mainhausen Die HAMEG Instruments GmbH bescheinigt die Konformität für das Produkt The HAMEG Instruments GmbH herewith declares conformity of the product HAMEG Instruments GmbH déclare la conformite du produit HAMEG
Content Content General information regarding the CE marking. . . . . . . 2 1 Installation and safety instructions. . . . . . . . . . . . 4 1.1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.2 Setting up the instrument. . . . . . . . . . . . . . . . . . . . . . . . 4 1.3 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.4 Intended Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.
Installation and safety instructions 1 Installation and safety instructions 1.1 Symbols (1) (2) (3) (4) (5) (6) (7) Symbol 1: Caution, general danger zone – Refer to product documentation Symbol 2: Risk of electric shock Symbol 3: Ground Symbol 4: PE terminal Symbol 5: ON/OFF supply voltage Symbol 6: Stand by display Symbol 7: Ground terminal 1.2 Setting up the instrument As you can see from the picture, there are small feets on the bottom which can be folded out.
Installation and safety instructions The measuring instrument is designed for use in the following sectors: ❙❙ Industrial sector ❙❙ Home ❙❙ Business and commercial sectors ❙❙ Small businesses. The measuring instrument is designed for indoor use only. Before each measurement, you need to verify at a known source if the measurement instrument functions properly. To disconnect from the mains, the rubber connector on the back panel has to be unplugged. 1.
Installation and safety instructions measurement input. The auxilary output (AUX OUT) is a multi-purpose output with the function as component tester, trigger output, pass/fail and function generator. When performing measurements in category II, III or IV circuits, it is mandatory to insert a probe that reduces the voltage so that no transient overvoltages will be applied to the instrument. Direct measurements (without galvanic isolation) to category II, III or IV circuits are prohibited.
Introduction 2 Introduction A 2.1 Front view On the instrument frontside you can find the power key 1 , in order to switch on the instrument or enter stand-bymode. If the instrument is in stand-by-mode, this key light up red. If the instrument is switched off using the main power switch on the backside, the red light will also switch off (this will take some seconds).
Introduction B 22 18 23 19 24 25 20 26 21 27 C 28 29 33 30 34 31 35 32 36 D 37 38 37 41 42 39 integrated HELP 16 and FILE/ PRINT 17 . Depending on how it is programmed, FILE/PRINT enables you to directly save instrument settings, signals or screen displays.
Introduction priate soft menu key. Once the last page was listed, the display will loop back to the the first page. 2.5 Basic Settings and Integrated Help You can access important basic settings such as language for user interface and help, general settings and interface settings in the menu that opens when you press the SETUP key in the GENERAL section. On the first page of the menu for the basic settings you can select the language for user interface and help.
Introduction Description HMO 1002 Options1) Upgrade Voucher2) Bandwidth upgrade from 50 MHz to 70 MHz HOO572 HV572 Bandwidth upgrade from 50 MHz to 100 MHz HOO512 HV512 Bandwidth upgrade from 70 MHz to 100 MHz HOO712 HV712 I2C, SPI, UART/RS-232 trigger and decoding option on analog and digital channels HOO10 HV110 I2C, SPI, UART/RS-232 trigger and decoding option on analog channels HOO11 HV111 CAN/LIN trigger and decoding option on analog and digital channels HOO12 HV112 Tab. 2.
Quick Start Guide 3 Quick Start Guide The following chapter is an introduction to the most important R&S®HMO1002 oscilloscope features and settings allowing you to use the instrument promptly. The integrated calibrator signal output is used as the signal source which means that the first steps will not require any additional instruments. 3.1 A Setting up and turning the instrument on 3 4 7 6 9 10 12 13 15 16 Fig. 3.3: Screen display after changing to DC coupling Fig. 3.
Quick Start Guide ❙❙ pos. pulse width, ❙❙ pos. duty cycle, neg. pulse width, neg. duty cycle. Fig. 3.6: ZOOM function 3.4 Cursor Measurements After displaying and reviewing the signal, the next step will introduce cursor measurements of the signal. Again, briefly press the AUTOSET key once followed by the CURSOR/MEASURE key. You can now select the desired type of measurement cursor from the menu. Press the soft menu key MEASURE TYPE to open the appropriate selection menu.
Quick Start Guide 3.6 Mathematical functions In addition to cursor and automatic measurements, your R&S®HMO1002 can also apply mathematical operations to the signals. Pressing the MATH 26 key opens a menu enabling you to apply addition, subtraction, multiplication or division to two analog channels. This also displays the mathematical graph. The top soft menu key allows you to select the first operand. With the key below, you can select the operator.
Vertical System 4 Vertical System You can select vertical settings using knobs for the vertical position and the 22 sensitivity. You may also use 18 a menu that is permanently 23 displayed as well as an ad19 vanced menu. You can select 24 for which channel you want the settings to be activated 25 by pressing the respective 20 channel key. Once a channel 26 has been activated, the key 21 will be marked by a colored 27 LED light. Additionally, the screen display will include a Fig. 4.
Vertical System SETUP key in the GENERAL section, and then press the soft menu key PROBE COMP. This compensation wizard will guide you through all the important points of the probe compensation. After connecting a probe, you can select the respective analog channel via soft menu key CHANNEL 1 (CH1) or CHANNEL 2 (CH2). The enclosed HZ154 probe stylus allows you to set the probe trimmers to an optimal square wave shape. The compensation is performed at 1 kHz (LF) and 1 MHz (HF).
Horizontal System 5 Horizontal System D 37 38 37 41 42 39 43 40 44 Fig. 5.1: Control panel of the horizontal system In the HORIZONTAL system section, users can select time base settings for capturing, trigger time position, zoom functions, acquisition modes, marker functions and search functions. Knobs allow you to set time base and trigger time position. A menu enables you to select the desired acquisition mode. A separate key is available to activate the zoom.
Horizontal System During peak detection, the oscilloscope distinguishes between two types: ❙❙ Acquisition peak detection: Each A/D converter converts at the full sampling rate (no Interlace mode), even if results have not been written to the acquisition memory at full sampling rate (for slow time bases, for instance). If peak detection is activated, unused converter values will be evaluated to detect minimum and maximum amplitudes.
Horizontal System window displays up to 40 detected data values (limited by processor performance). How much data is currently displayed depends on the displayed time window and the current sampling rate. If peak detection is activated, each The entire oscilloscope memory can only be read out in STOP mode if the maximum sampling rate has been activated. memory).
Horizontal System also possible to zoom out at the maximum repeat rate if the STOP mode was run at fast time bases. If a high repeat rate can only be attained by means of low memory depth (as is the case with other manufacturers) it is nearly impossible to zoom in retrospectively in STOP mode. 5.4 Interlace Mode In interlace mode, ADC and storage of two channels are connected. This doubles the sampling rate and the acquisition memory.
Horizontal System position and setting it to the center of the screen using the position knob, you can set an additional marker. This procedure allows you to mark up to 8 interesting positions within the signal. You can toggle between markers by 37 . These keys also allow you pressing the arrow keys to center the markers in the middle of the screen. This feature enables you to quickly compare marked signal sections in ZOOM mode.
Trigger System 6.3 Trigger type The trigger type can be selected with the TYPE key 17 in the TRIGGER section. A submenu will be opened. 6.3.1 Edge Trigger The easiest and by far the most frequently used trigger is the edge trigger. The oscilloscope triggers if slopes that were set with the EDGE key occur within the signal selected in the SOURCE menu. The signal slope has to pass through the set trigger level. The trigger type EDGE TRIGGER is selected in the Autosetup mode (AUTOSET key).
Trigger System MENU section. Selecting the corresponding soft menu item allows you to set any of these settings for positively (POS.) or negatively polarized pulses (NEG.). For the associated positive pulse, you define the width from rising to falling slopes, and accordingly for the associated negative pulse from falling to rising slopes. As is consistent with the principle, triggering always occurs on the second slope of the pulse. 6.3.
Trigger System 6.3.4 Video Trigger The video trigger allows you to trigger on PAL, NTSC SECAM standard video signals or on HDTV signals. After selecting the VIDEO trigger press the SOURCE 32 key. The FILTER 36 menu allows you to define additional settings. The oscilloscope triggers if the CVBS signal (Color Video Baseband Signal) selected in the SOURCE menu features the attributes set in the FILTER menu. Select the desired standard by pressing the respective soft menu key STANDARD.
Signal Display 7 Signal Display The following chapter describes the selection and display of signals from various sources as well as all available display modes. 7.1 Display Settings The R&S®HMO1002 features a high quality TFT display with VGA (640x480 pixels resolution) including LED backlighting. Basic display settings can be defined by pressing the DISPLAY 14 key in the GENERAL section. When the soft menu item VIRTUAL SCREEN is activated, a scroll bar will display to the right of the display graticule.
Signal Display 7.3 Signal Intensity Display and Persistence Function The default setting (indicated as active when the INTENS/ PERSIST 7 key is illuminated in white) allows you to use the universal knob to change the intensity of the signal display to anywhere from 0% to 100%. Persistence mode allows the display of varying signals by enabling the instrument to write several signals on the display simultaneously.
Signal Display 8 Measurements There are two different types of measurements on signals: cursor measurements and automatic measurements. All measurements are stored in a buffer memory that is larger than the display memory. The integrated hardware counter shows the frequency and period duration for the selected input. 8.1 Cursor Measurements The measurement option that is most frequently used with an oscilloscope is the cursor measurement. Depending on the measurement type up to three markers are available.
Measurements tes without voltage reference are displayed here. If “?” is displayed, the display does not show a complete measurement result. For instance, the period to be measured may not display completely and can consequently not be identified. PEAK + : This mode measures the maximum voltage value in the displayed view of the screen. The measurement will only be applied to the selected channel (Vp+). If the function AUTOM.
Measurements slope. This measurement type identifies the two slopes and calculates the pulse width from their time difference. The measurement will only be applied to the selected channel and requires a minimum of one completely displayed period of a triggered signal. . PULSE WIDTH –: This mode measures the width of the negative pulse. A negative pulse consists of a falling slope followed by a rising slope.
Analysis calculated from the amplitude of the measurement signal. A slope will be counted if the signal runs through the mean value. The measurement will only be applied to the selected channel. COUNT –/ : This mode counts signal changes (slopes) from High Level to Low Level in the displayed view of the screen. The mean value is calculated from the amplitude of the measurement signal. A slope will be counted if the signal runs through the mean value.
Analysis 9.2 Frequency Analysis (FFT) In general, the FFT in an oscilloscope works differently than in a spectrum analyzer and is affected not only by the time base setting, but also by the available number of used acquired data points when calculating the FFT. The R&S®HMO1002 allows you to include up to 128k point in the FFT. The FFT is not suitable for the analysis of very slow signals (Hzrange); this type of analysis requires a classic oscilloscope mode.
Analysis ased noise. This function can be used for pulse response tests with start and end values of zero. The menu item Y-SCALING allows you to scale the FFT in the amplitude logarithmically (dBm / dBV) or linear (Veff). The unit dBm (Decibel-Milliwatt) refers to 1 mW. The unit dBV (Decibel-Volt) refers to 1 Veff. The displayed values refer to a 50 Ohm terminating resistor. You can use an external terminating resistor parallel to the high impedance input.
Analysis ❙❙ Screen dump for first-time failure on USB stick ❙❙ Screen dump for first-time failure on printer An action is performed if the respective condition is met (e.g. a specific number of mask failures). Each action is assigned a unique condition which can be defined separately from the other actions. The respective condition can be defined in the menu for the corresponding action.
Analysis pected which will point to the defective component. Rather exact results may be achieved if the measurements are compared to those of intact components. This is especially true for semiconductors. The polarity of diodes or transistors can thus be identified if the lettering or marking is missing. Please note that with semiconductors changing the polarity will cause the display to rotate 180 degrees around the screen center.
Signal Generation 10 Signal Generation 10.1 Function Generator The R&S®HMO1002 includes an integrated function generator which can generate input signals while testing circuits, for instance. To activate the function generator, select the UTIL menu in the VERTICAL control panel, and then press the soft menu key FUNCTION GEN. The following waveforms can be generated and made available at the AUX OUT connector: ❙❙ DC ❙❙ SINE: Frequency range from 0.1 Hz to 50 kHz ❙❙ SQUARE WAVE: Frequency range from 0.
Signal Generation to a specific soft menu key and the condition HIGH (H) or LOW (L) is selected. 10.2.5 BUS Signal Source Use the pattern generator connectors on the front panel of the instrument to generate the following BUS signals (for measurements without measurement object), depending on the setting: ❙❙ SPI: Data rate 100 kBit/s, 250 kBit/s or 1 MBit/s ❙❙ I2C: Data rate 100 kBit/s, 400 kBit/s, 1 MBit/s or 3.4 MBit/s ❙❙ UART: Data rate 9600 Bit/s, 115.
Documentation, Storage and Recall 11 Documentation, Storage and Recall The oscilloscope allows all screen displays that store user settings (e.g. trigger condition and time base setting), reference curves and curves. An internal memory integrated with the instrument is available for reference curves and instrument settings. Screenshots and curve data can also be stored on a connected USB stick. The USB stick should not exceed 4 GB and must be FAT formatted (FAT32).
Documentation, Storage and Recall files by name, type, size or date. Press ACCEPT DIR. to confirm the target directory and you will automatically return to the curve main menu. The soft menu CURVE allows you to use the universal knob in the CURSOR/MENU section to select a channel which will be saved as a curve. You may only select channels that have been activated via channel keys. You can also save all visible channels simultaneously.
Documentation, Storage and Recall out the display memory or the entire acquisition memory. After you made all entries, press the menu key STORE to save the selected curve(s) according to the settings. 11.4 Screenshots The most important format to store information for documentation purposes is the screenshot. A screenshot is an image file which shows the current screen content at the time that storage takes place.
Mixed-Signal Operation 12 Mixed-Signal Operation As a standard, the R&S®HMO1002 is equipped with the connectors for the HO3508 logic probes to add 8 digital logic inputs. All software required for the mixed signal operation is already included in the firmware of each HMO. It is only necessary to purchase and connect the active HO3508 logic probes (8 channels). If you activate the POD, analog channel 2 will be deactivated automatically.
Mixed-Signal Operation can also use partially identical logic channels in the two possible buses. If you select PARALLEL CLOCKED as BUS TYPE, you can also use the bottom soft menu key CONTROL WIRES to select sources for CHIP SELECT, and you can use the universal knob to select the settings for CLOCK. The soft menu key ACTIVE is used to determine if the chip select signal High or Low Active is selected. The soft menu key SLOPE allows you to toggle between rising, falling and both slopes.
Serial Bus Analysis 13 Serial Bus Analysis 13.1 Software options (license key) The R&S®HMO1002 can be upgraded with options / voucher to trigger and decode serial buses. ❙❙ HOO10/HV110: The option HOO10 resp. the voucher HV110 can be used to trigger and decode I2C, SPI and UART/RS-232 buses on the digital channels (option logic probe HO3508 needed) and on the analog inputs. This option allows the decoding of two serial buses simultaneously. ❙❙ HOO11/HV111: The option HOO11 resp.
Serial Bus Analysis the bottom of the screen. Generally, a complete message of a protocol is displayed in a row. The columns include important information, e.g. address and date of the message. The number of rows in the table is identical to the number of complete message frames in storage. The decoding results may be saved as CSV file by using the soft menu key SAVE (e.g. save to a USB stick). Fig. 13.
Serial Bus Analysis ❙❙ Read/Write bit (R/W): Indicates, if the data is to be written or read out from the slave ❙❙ Acknowledge bit (ACK): Is issued by the recipient of the previous byte if transmission was successful (exception: for read access, the master terminates the data transmission with a NACK bit after the last byte) ❙❙ Data: a series of data bytes with a ACK bit after each byte ❙❙ Stop condition: rising slope on SDA (Serial Data), while SCL (Serial Clock) is HIGH 13.4.
Serial Bus Analysis for the trigger condition and the address. In most cases, the byte offset is zero if the trigger is to occur on the maximum first 24 bits after the address. The soft menu key NUMBER OF BYTES allows you to define how many bytes are to be analyzed for the trigger condition. The input may be binary or hexadecimal (PATTERN INPUT). If binary input is selected, the individual bits can be assigned to any condition via soft menu key SELECT BIT and the universal knob.
Serial Bus Analysis ❙❙ CLK: Data will be stored with rising or falling slope (rising slope is the default setting) ❙❙ DATA: Data high or low active (high active is the default setting) You can use the soft menu key BIT ORDER to determine if the data of each message starts with the MSB (most significant bit) or the LSB (least significant bit). The soft menu key WORD SIZE allows you to select via universal knob how many bits are included per message. You may select any value between 1 and 32 bits. button).
Serial Bus Analysis If the option HOO11 resp. the voucher HV111 is installed, it it only possible to select analog channels as source. If the option HOO10 resp. the voucher HV110 is installed, both analog and digital channels are available as source. Parity bits are used to detect errors during a transmission.
Serial Bus Analysis be marked with a green border in the display window of the trigger condition. ❙❙ PARITY ERROR: Trigger with a parity filter ❙❙ FRAME ERROR: Trigger with a frame error ❙❙ BREAK: Trigger with a break. The BREAK condition is fulfilled if a stop bit does not succeed a start bit within a specified time period. The stop bits low are active during the break. Press the MENU OFF button three times to close all menus, and the oscilloscope will trigger on the set bit sequence.
Serial Bus Analysis allows you to specify the data bit pattern or HEX values for up to 8 bytes (only available if DATA was selected as frame type). Available comparisons for address and data values are GREATER, EQUAL OR LESS, EQUAL and NOT EQUAL. Press the MENU OFF button twice or three times to close all menus, and the oscilloscope will trigger on the set data. For measurements without measuring object please refer to chapter 10.2.5. 13.
Serial Bus Analysis opens you can press the top soft menu key DATA to select the desired channel via universal knob in the CURSOR/ MENU section. The soft menu key POLARITY allows you to toggle between High and Low; the active function will be highlighted in blue. An analog or a digital channel can be connected to LIN-High or LIN-Low. You can select any version for the LIN standard (version 1x, version 2x, J2602 or any) via soft menu key VERSION and universal knob.
Serial Bus Analysis 14 Remote Control The R&S®HMO1002 is equipped with a built-in interface card, which have an Ethernet and USB connection on board as a standard. To make any communication possible, the chosen interface and it’s correcponding settings must be the same in the PC as in the oscilloscope.. In addition to a LAN interface, the R&S®HMO1002 includes a USB device port. For this interface, the user can select if the instrument is accessed via virtual COM port (VCP) or via USB TMC class. 14.
Remote Control Select “Next“ to start the installation and follow the installation instructions. Use the soft key to select “USB” and the soft menu PARAMETER. Fig. 14.2: NI-VISA installation instructions Fig. 14.5: Interface menu In this step, please select “NI-VISA xxx --> Leave this feature and its subfeatures installed locally“. Choose the USB type „USB TMC“. Fig. 14.6: Interface parameter menu Fig. 14.
Remote Control Once you open the Windows Device Manager, the following entry will be displayed: “USB Test and Measurement Devices --> USB Test and Measurement Device (IVI)”. Fig. 14.8: Device manager 14.3 Ethernet In addition to the USB interface, the interface card includes an Ethernet interface. Configure the settings in the oscilloscope for all necessary parameters after you select ETHERNET as interface and press the soft menu key PARAMETER.
Technical Data Technical DataData 15 Technical R&S®HMO1002 50/70/100MHz 2-channel mixed signal oscilloscopes Firmware: ≥ 5.0 All data valid at 23°C after 30 minute warm-up.
Technical Data Video logic channels pulse Sync. pulse polarity positive, negative supported standards NTSC, SECAM, PAL, PAL-M, SDTV 576i, HDTV 720p, HDTV 1080i, HDTV 1080p pre-trigger 0 to 500.
Technical Data Scale dBm, dBV, Vrms Waveform arithmetics refresh, envelope, average (up to 512) Cursor measurement 2 horizontal cursors, previous/next peak search Sources all analog channels reference waveforms: traces on external USB memory, available file formats: BIN (MSB/LSB), FLT (MSB/ LSB), CSV, TXT Pattern Generator Functions Square wave (Probe ADJ output) Bus Signal Source (4 Bit): square wave / probe adjust, bus signal source, counter, programmable pattern data display or acquisitio
Appendix 16 Appendix 16.1 List of figures Fig. 1.1: Fig. 1.2: Fig. 2.1: Fig. 2.2: Fig. 2.3: Fig. 3.1: Fig. 3.2: Fig. 3.3: Fig. 3.4: Fig. 3.5: Fig. 3.6: Fig. 3.7: Fig. 3.8: Fig. 3.9: Fig. 4.1: Fig. 4.2: Fig. 4.3: Fig. 5.1: Fig. 5.2: Fig. 5.3: Fig. 5.4: Fig. 5.5: Fig. 6.1: Fig. 6.2: Fig. 6.3: Fig. 6.4: Fig. 7.1: Fig. 7.2: Fig. 7.3: Fig. 9.1: Fig. 9.2: Fig. 9.3: Fig. 9.4: Fig. 9.5: Fig. 9.6: Fig. 10.1: Fig. 10.2: Fig. 11.1: Fig. 11.2: Fig. 11.3: Fig. 11.4: Fig. 12.1: Fig. 13.1: Fig. 13.2: Fig. 13.3: Fig. 13.
Appendix G Grayscale mode: 9 grid: 12, 15, 17, 19, 25, 39, 41 H Hamming window function: 30 Hanning window function: 30 Hold off time: 20 I I2C bus: 42, 43 interlace mode: 19 L licence key: 9, 10, 41 LIN bus: 48 Lissajous: 25 logic channels: 39, 40, 42 logic mode: 22 logic probe: 7, 8, 10, 20, 22, 39, 41 Logic Trigger: 22, 39 low pass filter: 14, 16, 21 M Maintenance: 5 mask test: 7, 31, 32 mathematical graph: 13 maximum repeat rate: 17, 18, 19 Maximum voltage: 31 mean value: 16, 26, 27, 28, 29, 30 mean vo
Appendix 58
Appendix 59
value-instruments.com value-instruments.com www.hameg.com HAMEG Instruments GmbH Industriestr. 6 | 63533 Mainhausen | Germany | Tel +49 (0) 6182 8000 R&S® is a registered trademark of Rohde & Schwarz GmbH & Co.