Datasheet
14 R&S®HMO1002
Accessories included:
Line cord, printed operating manual, 2 probes : HZ154 (100MHz, 10:1/1:1
switchable), HZ20 adapter BNC plug /4 mm banana socket, software-CD
software-CD
Mixed Signal Oscilloscope
50/70/100 MHz
R&S®HMO1002
Deutsch / Englisch
Benutzerhandbuch / User Manual
Benutzerhandbuch / User ManualUser Manual
Test & Measurement
Printed operating manual
HZ154
10/100 MHz
passive probe
HZ20
Adapter BNC plug /
4 mm banana socket
Window: Hanning, Hamming, Rectangular,
Blackman
Scale: dBm, dBV, V
rms
Waveform arithmetics: refresh, envelope, average (up to 512)
Cursor measurement: 2 horizontal cursors, previous/next peak
search
Sources: all analog channels
Pattern Generator
Functions: square wave / probe adjust, bus signal
source, counter, programmable pattern
Square wave
(Probe ADJ output):
frequency range: 1mHz to 500kHz
level: 2.5V
pp
(ta < 4ns)
polarity: normal, invert
duty cycle: 1% to 99%
Bus Signal Source (4 Bit):
I
2
C (100 kBit/s, 400 kBit/s, 1 MBit/s), SPI
(100 kBit/s, 250 kBit/s, 1 MBit/s), UART
(9600 Bit/s, 115,2 kBit/s, 1 MBit/s), CAN
(up to 50 MBits/s), LIN (up to 50 MBits/s)
Counter (4 Bit): frequency: 1 mHz to 25 MHz
direction: incrementing, decrementing
Programmable pattern (4 Bit):
sampling time: 20 ns to 42 s
memory depth: 2048 sa
pattern idle time: 20 ns to 42 s
Function Generator
Waveform modes: DC, sine, square, triangle/ramp, pulse
Sine: frequency range: 0.1 Hz to 50 kHz
flatness: ±0.5 dB relative to 1kHz
Square: frequency range: 0.1 Hz to 50 kHz
rise time: <4 µs
Triangle/ramp: frequency range: 0,1 Hz bis 10 kHz
Pulse: frequency range: 0.1 Hz to 10 kHz
duty cycle: 10% to 90%
Sampling rate: 978 kSa/s
Frequency accuracy: 50.0 x 10
-6
Aging: ±10.0 x 10
-6
per year
Amplitude
high impedance load: 60 mV
pp
to 6 V
pp
50 Ω load: 30 mV
pp
to 3 V
pp
Accuracy: 3%
DC offset: ±3 V
Digital Voltmeter
Display (3-digit):
Primary and secondary measurement
value per channel, simultaneous
measuring on all channels
Functions:
DC, DC
rms
, AC
rms
, V
pp
, V
p+
, V
p
-, crest factor
Sources all analog channels
Component Tester
Parameters: voltage (X), current (Y)
Testing frequency: 50 Hz, 200 Hz
Voltage: 10 V
p
(open)
Current: 10 mA (short)
Reference potential: Ground (PE)
Interfaces
for mass storage (FAT16/32): 1 x USB-Host (Typ A), max. 500 mA
for remote control: Ethernet (RJ45), USB Device (Typ B)
General Data
Application memory: 3 MB for references and device settings
Save/Recall
device settings:
on internal file system or external USB
memory, available file formats: SCP, HDS
reference waveforms:
on internal file system or external USB
memory, available file formats: BIN (MSB/
LSB), FLT (MSB/LSB), CSV, TXT, HRT
traces:
on external USB memory, available file
formats: BIN (MSB/LSB), FLT (MSB/LSB),
CSV, TXT
data: display or acquisition data
sources: single or all analog channels
screenshots: on external USB memory, available file
formats: BMP, GIF, PNG
Realtime Clock (RTC): date and time
Power supply:
AC supply: 100 V to 240 V, 50 Hz to 60 Hz, CAT-II
Power consumption: max. 25 W
Safety:
in line with IEC 61010-1 (ed. 3), IEC
61010-2-30 (ed. 1), EN 61010-1, EN
61010-2-030 , CAN/CSA-C22.2 No.
61010-1-12 , CAN/CSA-C22.2 No. 61010-
2-030-12 ,UL Std. No. 61010-1 (3rd
Edition) , UL61010-2-030
Operating temperature range: +5 °C to +40 °C
Storage temperature range: -20 °C to +70 °C
Rel. humidity: 5% to 80% (without condensation)
Mechanical Data
Dimensions (W x H x D): 285 mm x 175 mm x 140 mm
Net weight: 2.5 kg