Datasheet
12 R&S®HMO1002
R&S®HMO1002
50/70/100MHz 2-channel mixed signal oscilloscopes
Firmware: ≥ 5.0
All data valid at 23°C after 30 minute warm-up.
Display
Display: 16,5 cm (6,5“) VGA Color Display
Resolution: 640 (H) x 480 (V) Pixel
Backlight: 400 cd/m
2
(LED)
Display range in horizontal direction
without menu bar: 12 Div (600 Pixel)
with menu bar: 10 Div (500 Pixel)
Display range in vertical direction: 8 Div (400 Pixel)
with Virtual Screen usage: 20 Div
Color depth: 256 colors
Levels of trace brightness: 32
Trace display: pseudo-color, inverse brightnes
Vertical System
DSO Mode: CH1, CH2
MSO Mode: CH1, POD (with Logic Probe HO3508)
Analog channels
Y-bandwidth (-3dB)
(1 mV, 2 mV)/Div: 50 MHz
(5 mV to 10 V)/Div:
50 MHz (basic unit)
70 MHz (with HOO572/HV572 option)
100 MHz (with HOO512/HV512 or
HOO712/HV712 option)
Lower AC bandwidth: 2 Hz
Bandwidth limitation (switchable): about 20 MHz
Rise time (calculated, 10%-90%)
50 MHz (basic unit): <7 ns
basic unit with 70 MHz option: <5 ns
basic unit with 100 MHz option: <3.5 ns
DC gain accuracy (all ranges): 3% of full scale
Input sensitivity range
all analog channels: 1 mV/Div to 10 V/Div
coarse stepping: 13 calibrated steps, 1-2-5 sequence
variable stepping: freely between calibrated steps
Impedance: 1 MΩ II 16 pF ±2 pF
Coupling: DC, AC, GND
Max. input voltage: 200 V
p
(derates at 20 db/decade to 5 V
above 100 kHz)
Position range: ±5 Div (from center of screen)
Channel isolation: 35 dB from DC to specified bandwidth
(same V/Div range)
XY mode: CH1, CH2
Inversion: selectively all analog channels
Logic channels (with logic probe HO3508)
Thresholds: TTL, CMOS, ECL, user-definied (-2 V to
+8 V)
Impedance: 100 kΩ || 4 pF
Coupling: DC
Max. input voltage: 40 Vp
Trigger system
Trigger mode
Auto: Triggers automatically also without any
specific trigger event
Normal: Triggers only on specific trigger events
Single: Triggers once on a trigger event
Trigger indicator: Screen and panel (LED)
Trigger sensitivity
up to 5mV/Div: 1.5 Div
from 5mV/Div: 0.8 Div
Trigger level setting
with auto level: adjustable between peak values of a
signal
without auto level: ±5 Div (from center of screen)
external ±5.0 V
Trigger coupling
AC: <5 mV/Div: 10 Hz to 65 MHz
>5 mV/Div: 10 Hz to 65/90/130 MHz
DC: <5 mV/Div: DC to 65 MHz
>5 mV/Div: DC to 65/90/130 MHz
HF: <5 mV/Div: 30 kHz to 65 MHz
>5 mV/Div: 30 kHz to 65/90/130 MHz
selectable filters
LF
(low pass):
DC to 5 kHz (-3db), selectable in DC and
auto level mode
Noise rejection: min. level: 1.5 Div (> 5 mV/Div)
selectable with AC, DC and HF coupling
Trigger hold-off: auto, 50 ns to 10 s
External trigger input (BNC)
Impedance: 1 MΩ || 16 pF ±2 pF
Trigger level: 0.3 V
pp
to 10 V
pp
Max. input voltage: 100 V
p
Coupling: DC, AC
Trigger output via AUX OUT (BNC)
Functions:
Pulse output for every acquisition trigger
event, error output on mask violation
Output level: 3 V
Pulse polarity: positive
Pulse width: >150 ns (trigger event), >0.5 µs (mask
violation)
Trigger types
Edge
Direction rising, falling, both
Trigger coupling auto level AC, DC, HF
Switchable filters LF, noise rejection
Sources all analog and digital channels, AC line,
external (AC, DC)
Pulse width
Polarity positive, negative
Functions equal, not equal, lower, higher, within/
without a range
Pulse duration 16ns to 10s, resolution min. 2ns
Sources all analog channels
Logic
Functions:
Boolean operators: AND, OR, TRUE, FALSE
time based operators: equal, not equal, lower, higher, within/
without a time range, timeout
Duration: 16 ns to 10 s, resolution min. 2 ns
States: H, L, X
Sources: all logic channels