User manual

24
Horizontal System
time base (for a timebaseof100μs/Divthesmallesttime
is2μs,for1μs/Divthecorrespondingtimevalueis20ns).
If events match the search criterium, they will be highligh-
ted. The soft menu VIEW EVENT TABLE allows you to dis-
play the search results in a table format. Use the arrow
keys or the universal knob in the CURSOR/MENU section
to navigate the events in STOP mode. SET allows you to
center the selected event. When the zoom function is acti-
vated, the selected event will automatically be centered in
the zoom window.
Fig. 5.7: Search mode with event list
6 Trigger System
The trigger system of the
R&S®HMO Compact series
is easy to handle by just ob-
serving the concept of instru-
ment operation. There are 4
keys destined for frequently
used functions:
TYPE: selection of trigger
type (EDGE A/B), PULSE, LO-
GIC and VIDEO
SLOPE: type of slope
SOURCE: Determines the
triggers source
FILTER: Determines the
trigger conditions
Additional keys are available to select the trigger modes
(AUTO, NORMAL and SINGLE).
6.1 Trigger modes Auto, Normal, Single
The AUTO/NORM
29
key allows you to toggle directly
between the basic trigger modes. If Auto mode is activa-
ted, the key is not illuminated. Pressing the key will acti-
vate NORMAL mode, and a red LED will highlight the
key. In AUTO mode, the screen always displays a signal. If
asignalfulllsthetrigger conditions, the oscilloscope will
synchronize with this event and triggers when the set con-
ditionismet.Incaseofasignalthatdoesnotfulllthe
trigger condition (a simple case would be direct current),
the oscilloscope itself will generate a trigger event. This al-
lows a glance at the input signals at any time, regardless
of the trigger condition. In NORMAL mode, the signal will
now be captured and displayed if the trigger condition is
met.Incasenonewsignalfulllsthesettriggercondition,
the signal that was triggered last will be displayed. To en-
sure that only a signal that meets the trigger condition is
detected and displayed, press the Single key to
33
activate
this mode. This key is highlighted in white when the SIN-
GLE mode is activated. The R&S®HMO Compact series de-
tection and trigger system is now activated, indicated by
a blinking RUN/STOP key
39
. If the trigger condition is ful-
lled,thetriggersystemisactivated,dataisstoredandthe
oscilloscope switches to STOP mode (the RUN/STOP key
is permanently highlighted in red).
6.2 Trigger sources
The two resp. four analog channels are available as trigger
sources (SOURCE key). If the optional extension with ac-
tive logic probes R&S®HO3508 including 8 digital inputs
is connected, up to 8 digital inputs are available as trigger
source. The soft menu key AC LINE enables you to trigger
the trigger at system frequency. The trigger signal is ext-
racted internally from the power supply.