Datasheet

8 Rohde & Schwarz R&S®HMO3000 Series
Serial Bus Analysis
I
2
C, SPI, CAN or LIN – in terms of interaction with the outside world for embedded systems, it is safe to
say that these are the most commonly used communication protocols. The R&S®HMO3000 series by
Rohde & Schwarz offers you hardware-accelerated signal triggering and decoding for all of these
protocols. You can upgrade your instrument via software licence keys with those functions required to
develop your application:
R&S®HOO10 / R&S®HV110: Analysis of I
2
C, SPI and UART/RS-232 signals on analog and logic channels
R&S®HOO11 / R&S®HV111: Analysis of I
2
C, SPI and UART/RS-232 signals on all analog channels
R&S®HOO12 / R&S®HV112: Analysis of CAN and LIN signals on analog and logic channels
Serial bus trigger types:
I
2
C: Start, Stop, ACK, nACK,
Address/Data
SPI: Start, End, Serial Pattern (32Bit)
UART/RS-232: Startbit, Frame Start,
Symbol, Pattern
LIN: Frame Start, Wake Up,
Identier, Data, Error
CAN: Frame Start, Frame End,
Identier, Data, Error
SPI bus signal, MISO / MOSI decoded
HEX decoded CAN bus signal
I
2
C bus signal in zoom view
HMO3000_bro_de-en_3607-0146-3X_v0200.indd 8 29.02.2016 11:23:19