Datasheet
Rohde & Schwarz R&S®HMO3000 Series 13
Min. signal amplitude 0.8 Div
Trigger level range (seperately
adjustable with different
sources)
±8 Div (from center of screen)
external ±5.0 V
Trigger coupling
State A auto level, AC, DC, HF, LF, low-pass
State B
same sources as state A
different sources DC, HF, low-pass
Trigger setting
time based 16 ns to 8.589 s, resolution min. 4 ns
event based 1 to 2
16
events
Pulse Width
Polarity positive, negative
Functions equal, not equal, lower, higher, within/without
a range
Pulse duration 4 ns to 8.5 s, resolution min. 0.5 ns
Sources all analog and digital channels
Logic
Functions
boolean operators AND, OR, TRUE, FALSE
time based operators equal, not equal, lower, higher, within/without
a time range, timeout
Duration 4 ns to 8.5 s, resolution min. 0.5 ns
States H, L, X
Sources all logic channels
Video
Sync. pulse polarity positive, negative
Supported standards NTSC, SECAM, PAL, PAL-M, SDTV 576i, HDTV
720p, HDTV 1080i, HDTV 1080p
Field even/odd, either
Line line number selectable, all
Sources all analog channels, external (AC, DC)
Risetime
Functions rise/fall time, both
Time range 4 ns to 8.5 s, resolution min. 0.5 ns
Time based operators equal, not equal, lower, higher
Variance ±2 ns to ±33.5 ms, resolution 2 ns
Sources all analog channels
Runt
Polarity positive, negative, both
Sources all analog channels
Serial Busses (optional)
Bus representation
Up to two busses can be analyzed at the same
time. Color-coded display of decoded data in
ASCII, binary, decimal and hexadecimal format.
Option / Voucher codes
R&S®HOO10 / R&S®HV110 Analysis of I
2
C, SPI, UART/RS-232 signals on
analog and logic channels
R&S®HOO11 / R&S®HV111 Analysis of I
2
C, SPI, UART/RS-232 signals on
all analog channels
R&S®HOO12 / R&S®HV112 Analysis of CAN and LIN signals on analog and
logic channels
Trigger types by protocols
I
2
C Start, Stop, ACK, NACK, Address/Data
SPI Start, End, Serial Pattern (32 Bit)
UART/RS-232 Startbit, Frame Start, Symbol, Pattern
LIN Frame Start, Wake Up, Identifier, Data, Error
CAN Frame Start, Frame End, Identifier, Data, Error
Horizontal System
Time domain (Yt) main screen, time domain and zoom window
Frequency domain (FFT) time domain and frequency domain window
(FFT)
XY/XYZ mode voltage (XY), intensity (Z)
VirtualScreen virtual display of 20 Div for all math, logic, bus,
reference signals
Reference signals up to 4 references
Channel deskew -62.5 ns to +61.5 ns, step size 500 ps
Memory zoom up to 250,000:1
Time basis
accuracy ±15.0 x 10
-6
aging ±5.0 x 10
-6
per year
Operation modes
REFRESH 1 ns/Div to 50 s/Div
ROLL 50 ms/Div to 50 s/Div
Acquisition System
Realtime Sampling Rate
2-channel models 2 x 2 GSa/s or 1 x 4 GSa/s
4-channel models 4 x 2 GSa/s or 2 x 4 GSa/s
logic channels 16 x 1 GSa/s
Memory Depth
2-channel models 2 x 4 MPts or 1 x 8 MPts
4-channel models 4 x 4 MPts or 2 x 8 MPts
Resolution 8 bit, (HiRes up to 16 bit)
Waveform arithmetics
refresh, roll (loose/triggered), average (up to 1024),
envelope, peak detect (500 ps), filter (low-pass,
adjustable), high resolution (up to
16 bit
)
Record modes
automatic, max. sampling rate, max. waveform
update rate, specific record length (10 kPts to
2 MPts)
Interpolation
all analog channels sin(x)/x, linear, sample-hold
logic channels pulse
Delay
pre-trigger 0 to 4x10
6
Sa x (1/sample rate), x2 in interlaced
mode
post-trigger 0 to 8,59 x 10
9
Sa x (1/sample rate)
Waveform update rate up to 5000 Wfm/s
Waveform display dots, vectors, persistence afterglow
Persistence afterglow min. 50 ms
Segmented Memory (optional, R&S®HOO14 / R&S®HV114)
Segment size 5 kPts to 1 MPts
max. number of segments up to 1,000
Re-arming time less than 3 µs
Sampling rate 200.000 Wfm/s
Segment player
Displays all recorded segments manually or
automatically. All measurement functions
including pass/fail testing can be applied on
the recorded segments.
Sources all analog and digital channels, busses
Waveform Measurements and Operation
Operation menu-driven (multilingual), auto-set, help
functions (multilingual)
Automatic measurements
voltage (V
pp
, V
p+
, V
p-
, V
rms
, V
avg
, V
min
, V
max
),
amplitude, phase, frequency, period, rise/fall
time (80%, 90%), overshoot (pos/neg), pulse
width (pos/neg), burst width, duty cycle (pos/
neg), standard deviation, delay, crest factor,
edge/pulse count (pos/neg), trigger period,
trigger frequency
Automatic search functions edge, pulse, peak, rise/fall time, runt
HMO3000_bro_de-en_3607-0146-3X_v0200.indd 13 29.02.2016 11:23:24