Scope Rider RTH
6Getting Started 1326.1561.02 ─ 0301
Input channels
●
R&S RTH1002: 2 oscilloscope channels, 1 multimeter
●
R&S RTH1004: 4 oscilloscope channels
Maximum input voltage
●
At BNC inputs: CAT IV 300 V
●
With probe R&S RT-ZI10 or R&S RT-ZI11: CAT IV 600 V, CAT III 1000 V
●
Meter input: CAT IV 600 V; CAT III 1000 V
Voltage ratings are V RMS (50 to 60 Hz) for AC sine wave and V DC for DC
applications.
Vertical system
●
Analog bandwidth (–3 dB) of oscilloscope channels:
–
R&S RTH1002 and R&S RTH1004: ≥ 60 MHz
– R&S RTH1002 with -B221 option and R&S RTH1004 with -B241 option:
≥ 100 MHz
– R&S RTH1002 with -B222 option and R&S RTH1004 with -B242 option:
≥ 200 MHz
– R&S RTH1002 with -B223 option and R&S RTH1004 with -B243 option:
≥ 350 MHz
– R&S RTH1002 with -B224 option and R&S RTH1004 with -B244 option:
≥ 500 MHz
●
Rise time (calculated):
R&S RTH1002 and R&S RTH1004 with bandwidth of 60 MHz, 100 MHz,
200 MHz, 350 MHz, 500 MHz: < 5.8 ns, < 3.5 ns, < 1.75 ns, < 1 ns, < 700 ps
respectively
●
DC gain accuracy:
Offset and position set to zero after self-alignment
– Input sensitivity > 5 mV/div: ±1.5 %
– Input sensitivity > 2 mV/div to 5 mV/div : ±2 %
– input sensitivity 2 mV/div: ±2,5 %
●
Input impedance: 1 MΩ ± 1 % || 12 pF ± 2 pF (meas.)
●
Input coupling: DC, AC
●
Input sensitivity: 2 mV/div to 100 V/div
Main Specifications