491Bedienhandbuch 1326.1578.03 ─ 10
LOGic:GROup<m>:THReshold?
Returns the current threshold value.
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
Rückgabewerte:
<Level> Bereich: -10 bis 10
Inkrement: 1E-3
*RST: 0
Std-einheit: V
Verwendung: Nur Abfrage
LOGic:GROup<m>:HYSTeresis
<Hysteresis>
Hysteresis avoids the change of signal states due to noise oscillation around the thres-
hold level. Set a small hysteresis for clean signals, and large hysteresis for noisy sig-
nals.
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if LOGic:THCoupling is OFF.
Parameter:
<Hysteresis> SMALl | MEDium | LARGe
*RST: MEDium
LOGic:CHANnel<m>:DESKew <Value>
Sets the deskew for all channels of a logic probe at once, or for each logic channel
separately.
Suffix:
<m>
.
1..9
1..8: logic channels 0 to 7
9: all logic channels
Parameter:
<Value> Bereich: -100E-9 bis 100E-9
Inkrement: 800E-12
*RST: 0
Std-einheit: s
Logikanalysator (R&S RTH-B1 MSO)