490Bedienhandbuch 1326.1578.03 ─ 10
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if LOGic:THCoupling is OFF.
Parameter:
<ThresholdType> TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
TTL
1.4 V
ECL
-1.3 V
CMOS
2.5 V
GND
0 V (for CAN channels, requires option R&S RTH-K3)
CAN
2 V (for CAN channels, requires option R&S RTH-K3)
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V (for LIN channels, requires option R&S RTH-
K3)
12 V Supply
12 V (for LIN channels, requires option R&S RTH-K3)
18 V Supply
18 V (for LIN channels, requires option R&S RTH-K3)
USER
Set the value with LOGic:GROup<m>:USER.
*RST:
TTL
LOGic:GROup<m>:USER <ThresholdValue>
Sets the threshold value if LOGic:GROup<m>:TECHnology is set to USER.
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if LOGic:THCoupling is OFF.
Parameter:
<ThresholdValue> Bereich: -400 bis 400
Inkrement: 1E-3
*RST: 1.4
Std-einheit: V
Logikanalysator (R&S RTH-B1 MSO)