Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
482Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<Relation> UNUSed | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
*RST: EQUal
TRIGger:SENT:IDENtifier <Pattern>
Specifies the identifier pattern to be found, in binary format. Enter the pattern in MSB
first bit order.
Parameter:
<Pattern> String with max. 7 characters. Characters 0, 1, and X are allo-
wed. If you define a pattern shorter than the pattern length, the
missing LSB are filled with X.
TRIGger:SENT:SCONdition
<Relation>
Defines how the specified status pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameter:
<Relation> UNUSed | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
*RST: EQUal
TRIGger:SENT:STATus <Pattern>
Defines the status pattern as trigger condition. Enter the words in MSB first bit order.
Parameter:
<Pattern>
TRIGger:SENT:FCRCerror <FastCRCError>
Triggers on a detected CRC error in the transmission sequences of the fast channel.
The CRC length is 4 bits.
Parameter:
<FastCRCError> ON | OFF
*RST: ON
TRIGger:SENT:IRFLength <FrameLengthError>
Triggers on frame length errors in transmission sequences, if BUS:SENT:PPULse is
set to PPFL. A frame length error occurs, when the total length of the transmission
sequence (including pause pulse) does not match the frame length set with BUS:
SENT:PPFLength.
Protokollanalyse