479Bedienhandbuch 1326.1578.03 ─ 10
BUS:SENT:DATA:SOURce <Channel>
Sets the source of the data line. All channel waveforms can be used.
Parameter:
<Channel> C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
C3 and C4: only R&S RTH1004
Digital channels are available if option R&S RTH-B1 is installed.
*RST: C1
Verwendung: Asynchroner Befehl
BUS:SENT:DATA:THReshold
<ThresholdValue>
Sets the threshold value for digitization of signals manually. If the signal value is higher
than the threshold, the signal state is high. Otherwise, the signal state is considered
low.
This value is only considered for BUS:SENT:TECHnology auf Seite 480 USER.
Parameter:
<ThresholdValue> Bereich: -400 bis 400
Inkrement: 1E-3
*RST: 1.4
Std-einheit: V
BUS:SENT:DNIBbles <DataNibbles>
Sets the number of data nibbles for a transmission sequence.
Parameter:
<DataNibbles> Bereich: 1 bis 6
Inkrement: 1
*RST: 3
BUS:SENT:PPFLength <FrameLength>
Defines a constant transmission sequence length. To select the fixed sequence length,
set BUS:SENT:PPULse to PPFL.
Parameter:
<FrameLength> Bereich: 104 bis 922
Inkrement: 1
*RST: 128
BUS:SENT:PPULse <PausePulse>
Determines if a pause pulse is part of the SENT transmission sequence.