Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
478Bedienhandbuch 1326.1578.03 ─ 10
BUS:SENT:DNIBbles..................................................................................................... 479
BUS:SENT:PPFLength...................................................................................................479
BUS:SENT:PPULse....................................................................................................... 479
BUS:SENT:SFORmat.....................................................................................................480
BUS:SENT:TECHnology.................................................................................................480
BUS:SENT:RDSL...........................................................................................................480
BUS:SENT:CLKPeriod <ClockPeriod>
Sets the nominal clock period (clock tick).
Parameter:
<ClockPeriod> Bereich: 1E-6 bis 100E-6
Inkrement: 1E-6
*RST: 3E-6
Std-einheit: s
BUS:SENT:CLKTolerance <ClockTolerance>
Sets a tolerated deviation of the clock signal.
Parameter:
<ClockTolerance> Bereich: 0 bis 25
Inkrement: 1
*RST: 20
Std-einheit: %
BUS:SENT:CRCMethod <CRCCalculation>
Selects the calculation method for the CRC checksum.
Parameter:
<CRCCalculation> SAEJ | TLE
SAEJ: according to the standard
TLE: according to the computing method for TLE_4998X sen-
sors.
*RST: SAEJ
BUS:SENT:CRCVersion <CRCVersion>
Selects the calculaction method for the cyclic redundancy check (CRC).
Parameter:
<CRCVersion> V2010 | LEGA
LEGAcy: method used up to 2010
V2010: current method
*RST: V2010
Protokollanalyse