472Bedienhandbuch 1326.1578.03 ─ 10
Beispiel:
TRIG:LIN:TYPE ID
TRIG:LIN:IDEN 001001
TRIG:LIN:ICON EQU
Triggers if the ID of the measured signal is 001001.
TRIGger:LIN:IPERror <ParityError>
Triggers on a parity error if TRIGger:LIN:TYPE is set to ERRC
Parameter:
<ParityError> ON | OFF
*RST: ON
TRIGger:LIN:SYERror
<SyncError>
Triggers on a synchronization error if TRIGger:LIN:TYPE is set to ERRC
Parameter:
<SyncError> ON | OFF
*RST: ON
TRIGger:LIN:TYPE <TriggerType>
Parameter:
<TriggerType> SYNC | WKFR | ERRC | ID | IDDT
SYNC
Triggers on the stop bit of the sync field.
WKFR
Triggers after a wakeup frame.
ERRC
Identifies various errors in the frame. You can select one or
more error types as the trigger condition.
ID
Sets the trigger to a specific identifier or an identifier range.
IDDT
Sets the trigger to a combination of identifier and data condition.
The instrument triggers at the end of the last byte of the speci-
fied data pattern.
*RST:
SYNC
15.11.6.3 LIN-Decodierungsergebnisse
Das Framesuffix <m> gibt den Frameindex an, für den das Ergebnis abgefragt wird.
Das Bytesuffix <n> gibt den Byteindex an, für den das Ergebnis abgefragt wird.
Protokollanalyse