Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
470Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<ThresholdValue> Bereich: -400 bis 400
Inkrement: 1E-3
*RST: 1.4
Std-einheit: V
BUS:LIN:TECHnology <ThresholdType>
Sets the threshold value for digitization of signals according to the specified techno-
logy. If the signal value is higher than the threshold, the signal state is high. Otherwise,
the signal state is considered low.
To set a user-defined threshold, select USER and define the value using BUS:LIN:
DATA:THReshold auf Seite 469.
Parameter:
<ThresholdType> TTL | ECL | CMOS | CAN | GND | LIN7vsupply | LIN12vsupply |
LIN18vsupply | USER
CMOS
2.5 V
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V
USER
Set the value with BUS:LIN:DATA:THReshold.
*RST: TTL
BUS:SETReflevels
Sets the appropriate threshold or threshold for the selected serial protocol.
Verwendung:
Ereignis
15.11.6.2 LIN-Trigger
TRIGger:LIN:CHKSerror................................................................................................. 470
TRIGger:LIN:DATA........................................................................................................ 471
TRIGger:LIN:DCONdition................................................................................................471
TRIGger:LIN:ICONdition................................................................................................. 471
TRIGger:LIN:IDENtifier................................................................................................... 471
TRIGger:LIN:IPERror..................................................................................................... 472
TRIGger:LIN:SYERror.................................................................................................... 472
TRIGger:LIN:TYPE.........................................................................................................472
TRIGger:LIN:CHKSerror <ChecksumError>
Triggers on a checksum error if TRIGger:LIN:TYPE is set to ERRC
Protokollanalyse