Beispiel:
LOGGer:MEASurement1:VERTical:AUTO 0 //sets manual scaling for ch1
LOGGer:MEASurement1:VERTical:AUTO //autoscales logger channel 1
LOGGer:MEASurement1:VERTical:AUTO?
<-- 0 //state is unchanged (off)
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:UPPer <Upper>, [<Lower>]
LOGGer:MEASurement<m>:VERTical:LOWer <Lower>, [<Upper>]
Set the upper and lower range limits for the specified logger channel. The values must
differ. You can set both values using only one of the commands (see examples). The
query returns the mandatory parameter.
The commands disable the vertical auto scaling.
Suffix:
<m>
.
1..4
Logger channel
Parameter:
<Lower> Lower range limit
Einstellparameter:
<Upper> Upper range limit
Beispiel: Set the lower value to -1 and the upper value to +1 using both
commands:
LOGGer:MEASurement1:VERTical:LOWer -1
LOGGer:MEASurement1:VERTical:UPPer +1
Beispiel: Set the lower value to -1 and the upper value to +1 using one
command:
LOGGer:MEASurement1:VERTical:LOWer -1, +1
//or
LOGGer:MEASurement1:VERTical:UPPer +1, -1
Firmware/Software: FW 1.60 and higher
LOGGer:MEASurement<m>:VERTical:DEViation <Deviation>
Sets the distance between the lower range and the mean value and between the upper
range and the mean value.
The command disables the vertical auto scaling.
See also: LOGGer:MEASurement<m>:VERTical:MEAN.
Suffix:
<m>
.
1..4
Logger channel
Daten-Logging