Fernsteuerbefehle
R&S
®
Scope Rider RTH
416Bedienhandbuch 1326.1578.03 ─ 10
LOGGer:SOURce <Source>
Sets the logger source.
Befor you can change the logger source, stop the measurement.
Parameter:
<Source> SCOPe | METer | COUNter
SCOPe: an active automatic measurement is required.
METer | COUNter: The selected measurement must be active,
but not running.
LOGGer:TIMebase:SRATe <NextSampleRate>
Sets the number of samples per second.
Parameter:
<NextSampleRate> SA1 | SA2 | SA5 | SA10
*RST: SA1
LOGGer:TIMebase:SCALe <NextHorizScale>
Selects the horizontal scale of the logged data.
Parameter:
<NextHorizScale> AUTO | S1 | S2 | S4 | S5 | S10 | S20 | S40 | M1 | M2 | M4 | M5 |
M10 | M20 | M40 | H1 | H2 | H4 | H5 | H10 | H20 | D1 | D2 | D4
S1 | S2 | S4 | S5 | S10 | S20 | S40
Seconds per division
M1 | M2 | M4 | M5 | M10 | M20 | M40
Minutes per division
H1 | H2 | H4 | H5 | H10 | H20
Hours per division
D1 | D2 | D4
Days per division
*RST: AUTO
LOGGer:SLOT:CURRent <SelectedSlot>
Selects one of the 10 memory slots to store the data during the logging. It is possible to
change the slot during recording. This command shows the slot number that is selec-
ted in the menu.
Parameter:
<SelectedSlot> SLOT1 | SLOT2 | SLOT3 | SLOT4 | SLOT5 | SLOT6 | SLOT7 |
SLOT8 | SLOT9 | SLOT10
*RST: SLOT1