Fernsteuerbefehle
R&S
®
Scope Rider RTH
413Bedienhandbuch 1326.1578.03 ─ 10
Parameter:
<Range> <numeric value> | MIN | MAX | DEF
Defines the measurement range of the input signal for the coun-
ter. If probes are connected, the MIN | MAX | DEF values are
adapted accordingly.
<numeric value>
Power range in V. If necessary, the next higher range is selec-
ted.
MIN
100 mV
MAX
300 V
DEF
100 mV
Rückgabewerte:
<Counter result> Measured frequency
Verwendung: Nur Abfrage
COUNter<m>:READ?
Starts a new measurement, returns the current measurement result and stays in run
mode.
Suffix:
<m>
.
1..2
Rückgabewerte:
<Counter result> Measured frequency
Verwendung: Nur Abfrage
COUNter<m>:SENSe:FUNCtion [<Function>]
Selects the input channel for the counter.
Suffix:
<m>
.
1..2
Parameter:
<Function> 'FREQ 1' | 'FREQ 2' | 'FREQ 3' | 'FREQ 4'
Depends on available channels.
COUNter<m>:SENSe:RANGe <MeterRangeUI>
This command defines the measurement range.
Suffix:
<m>
.
1..2
Zähler-Modus (R&S RTH-K33)