Fernsteuerbefehle
R&S
®
Scope Rider RTH
412Bedienhandbuch 1326.1578.03 ─ 10
Beispiel:
COUN1:CONF:VAL?
//Result: 'FREQ 3V'
Verwendung: Nur Abfrage
COUNter<m>:FETCh?
Queries the currently measured value.
Suffix:
<m>
.
1..2
Rückgabewerte:
<ResultActual> Measured value
Bereich: -100E+24 bis 100E+24
Inkrement: 0
*RST: 0
Std-einheit: Hz
Verwendung: Nur Abfrage
COUNter<m>:INITiate
Resets the hardware and all statistical values, performs a single measurement and
ends in hold mode.
Suffix:
<m>
.
1..2
Verwendung: Ereignis
COUNter<m>:MEASure:FREQuency? <Range>
This command sets all measurement parameters and trigger parameters to default
values for frequency measurements, then immediately triggers a counter measurement
and returns the result.
This command corresponds to the commands COUNter<m>:CONFigure:
FREQuency + COUNter<m>:FETCh?
Suffix:
<m>
.
1..2
Zähler-Modus (R&S RTH-K33)