Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
411Bedienhandbuch 1326.1578.03 ─ 10
Rückgabewerte:
<Minimum> Average measured value
Bereich: -100E+24 bis 100E+24
Inkrement: 1E-12
*RST: 0
Std-einheit: Hz
Verwendung: Nur Abfrage
COUNter<m>:CONFigure:FREQuency <Range>
This command sets all measurement parameters and trigger parameters to default
values for frequency measurements.
Note: this command can be executed at any time, even if the counter is not yet activa-
ted. The parameters are stored internally and used when the counter is activated.
Suffix:
<m>
.
1..2
Parameter:
<Range> <numeric value> | MIN | MAX | DEF
Defines the measurement range of the input signal for the coun-
ter. If probes are connected, the MIN | MAX | DEF values are
adapted accordingly.
<numeric value>
Power range in V. If necessary, the next higher range is selec-
ted.
MIN
100 mV
MAX
300 V
DEF
100 mV
Verwendung:
Nur Einstellung
COUNter<m>:CONFigure:VALue?
Returns the current configuration for the counter measurement defined by the most
recent COUNter<m>:CONFigure:FREQuency or COUNter<m>:MEASure:
FREQuency? command.
Suffix:
<m>
.
1..2
Rückgabewerte:
<ConfigString> String containing the current measurement function (FREQ) and
the defined measurement range of the used input channel.
Zähler-Modus (R&S RTH-K33)