Instructions

Table Of Contents
Fernsteuerbefehle
R&S
®
Scope Rider RTH
405Bedienhandbuch 1326.1578.03 ─ 10
15.8.5 Ergebnisse
Werte lesen...........................................................................................................405
METer:MEASure-Befehle......................................................................................406
Statistik..................................................................................................................407
15.8.5.1 Werte lesen
METer<m>:INITiate........................................................................................................405
METer<m>:READ?........................................................................................................ 405
METer<m>:FETCh?....................................................................................................... 405
METer<m>:INITiate
Resets the hardware and all statistical values, performs a single measurement and
ends in hold mode.
Suffix:
<m>
.
1..4
Verwendung: Ereignis
METer<m>:READ?
Starts a new measurement, returns the current measurement result and stays in run
mode.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Rückgabewerte:
<Meter result>
Verwendung: Nur Abfrage
METer<m>:FETCh?
Returns the currently measured value.
Use this command after METer<m>:INITiate.
Suffix:
<m>
.
1..4
R&S RTH1002: always 1, omit the suffix
R&S RTH1004: Selects the voltmeter.
Voltmeter (R&S RTH1004)